Suppr超能文献

由于视光学部出现钙沉积导致硅酮人工晶状体混浊。

Opacification of a silicone intraocular lens caused by calcium deposits on the optic.

作者信息

Wackernagel Werner, Ettinger Karl, Weitgasser Ursula, Bakir Berin Gŭl, Schmut Otto, Goessler Walter, Faschinger Christoph

机构信息

Department of Ophthalmology, Karl-Franzens University Graz, Graz, Austria.

出版信息

J Cataract Refract Surg. 2004 Feb;30(2):517-20. doi: 10.1016/S0886-3350(03)00672-2.

Abstract

We describe opacification of a plate-haptic silicone intraocular lens (IOL) caused by calcification in a diabetic patient with asteroid hyalosis. The IOL was explanted 48 months after uneventful phacoemulsification because opacification of the posterior surface was causing significant visual disturbance. Light and scanning electron microscopy and x-ray spectrometry of the explanted IOL showed the opacification consisted mainly of calcium and phosphate, presumably hydroxyapatite, in the form of precipitations on the posterior surface of the optic.

摘要

我们描述了一名患有星状玻璃体病变的糖尿病患者,其植入的一片式触觉硅酮人工晶状体(IOL)因钙化而发生混浊。在白内障超声乳化手术顺利完成48个月后,该IOL被取出,因为其后表面混浊导致了严重的视觉障碍。对取出的IOL进行光镜、扫描电子显微镜检查及X射线光谱分析显示,混浊主要由钙和磷酸盐组成,推测为羟基磷灰石,以沉淀物的形式存在于晶状体光学部的后表面。