Certosimo Fred J, Diefenderfer Kim E, Mosur Michael A
Virginia Commonwealth University, School of Dentistry, Richmond, USA.
Gen Dent. 2003 Mar-Apr;51(2):142-6.
Following a pilot study which determined that autoclaving a disposable tip resulted in physical deformation, rendering the tip ineffective for photocuring, this study evaluated the effects of two FDA-cleared high-level disinfectants on the light output and physical integrity of disposable and nondisposable curing light tips. Nine disposable and five nondisposable curing tips were immersed in each of the two disinfectants. No significant differences (p > 0.05) in light output over time were observed for either type of curing light tip in either solution. None of the tips were physically affected by immersion and light output values remained in the acceptable range for resin composite polymerization.
在一项初步研究确定对一次性光固化头进行高压灭菌会导致物理变形,使光固化头无法用于光固化后,本研究评估了两种经美国食品药品监督管理局(FDA)批准的高水平消毒剂对一次性和非一次性光固化头的光输出及物理完整性的影响。将九个一次性光固化头和五个非一次性光固化头分别浸入两种消毒剂中。在两种溶液中,两种类型的光固化头随时间的光输出均未观察到显著差异(p>0.05)。没有一个光固化头因浸泡而受到物理影响,并且光输出值仍在树脂复合材料聚合的可接受范围内。