Br Dent J. 2011 Apr 23;210(8):E12. doi: 10.1038/sj.bdj.2011.312.
This study investigated the effects that disposable infection control barriers and physical damage through use had on the power output from dental light curing units (LCUs) and light curing tips (LCTs). Five disposable infection control barriers were tested on a number of LCUs and LCTs. Testing involved the repeated measurement of power output of LCUs and LCTs on a radiometer. Two of the barriers tested caused statistically significant reductions in the mean light output intensity when compared to the no barrier control groups. One barrier type reduced the power output by 30 to 40%. It was also noted that physical damage to the LCTs affected power output by between 20 and 30%, which was then further reduced by the disposable barrier. This study showed that three of the five disposable infection control barriers had little effect on the overall efficiency of the power output of the LCUs. It also showed that physical damage to LCUs and LCTs can affect power output significantly. Infection control measures should be carefully considered before use to avoid undue effects on power output delivered from the LCUs/LCTs to ensure that the degree of polymerisation within the resin-based composite and curing efficiency are not affected unduly.
本研究调查了一次性感染控制屏障和使用造成的物理损坏对牙科光固化机(LCU)和光固化尖(LCT)的功率输出的影响。对多个 LCU 和 LCT 测试了五种一次性感染控制屏障。测试涉及使用辐射计对 LCU 和 LCT 的功率输出进行重复测量。与无屏障对照组相比,两种测试的屏障类型导致平均光输出强度的统计学显着降低。一种屏障类型将功率输出降低了 30%至 40%。还注意到,LCT 的物理损坏会使功率输出降低 20%至 30%,然后再被一次性屏障进一步降低。本研究表明,五种一次性感染控制屏障中有三种对 LCU 功率输出的整体效率影响不大。它还表明,LCU 和 LCT 的物理损坏会显着影响功率输出。在使用之前,应仔细考虑感染控制措施,以避免对 LCU/LCT 输送的功率输出产生不必要的影响,以确保树脂基复合材料的聚合程度和固化效率不受过度影响。