Use of optical coherence tomography to monitor biological tissue freezing during cryosurgery.

作者信息

Choi Bernard, Milner Thomas E, Kim Jeehyun, Goodman Jared N, Vargas Gracie, Aguilar Guillermo, Nelson J Stuart

机构信息

Beckman Laser Institute, University of California, Irvine, Irvine, California 92612, USA.

出版信息

J Biomed Opt. 2004 Mar-Apr;9(2):282-6. doi: 10.1117/1.1648647.

Abstract

The use of optical coherence tomography (OCT) for imaging skin during cryosurgery is evaluated. OCT provides high spatial resolution (5-10 microm) images of optical backscattering due to local variations in refractive index, such as the boundary between liquid and frozen water in tissue. Time resolved OCT images were acquired during freezing of water, Intralipid trade mark, and in vivo hamster skin. Subsurface morphological changes were evident only during freezing of Intralipid and skin. A simple thermal model was applied which predicted freezing times on the same order of magnitude as those observed in OCT images. OCT can be used as a feedback tool during cryosurgical procedures to monitor progression of the freezing front.