Huda Walter, Lieberman Kristin A, Chang Jack, Roskopf Marsha L
Department of Radiology, SUNY Upstate Medical University, Syracuse, New York 13210, USA.
Med Phys. 2004 Mar;31(3):595-601. doi: 10.1118/1.1646233.
We investigated how patient head characteristics, as well as the choice of x-ray technique factors, affect lesion contrast and noise values in computed tomography (CT) images. Head sizes and mean Hounsfield unit (HU) values were obtained from head CT images for five classes of patients ranging from the newborn to adults. X-ray spectra with tube voltages ranging from 80 to 140 kV were used to compute the average photon energy, and energy fluence, transmitted through the heads of patients of varying size. Image contrast, and the corresponding contrast to noise ratios (CNRs), were determined for lesions of fat, muscle, and iodine relative to a uniform water background. Maintaining a constant image CNR for each lesion, the patient energy imparted was also computed to identify the x-ray tube voltage that minimized the radiation dose. For adults, increasing the tube voltage from 80 to 140 kV changed the iodine HU from 2.62 x 10(5) to 1.27 x 10(5), the fat HU from -138 to -108, and the muscle HU from 37.1 to 33.0. Increasing the x-ray tube voltage from 80 to 140 kV increased the percentage energy fluence transmission by up to a factor of 2. For a fixed x-ray tube voltage, the percentage transmitted energy fluence in adults was more than a factor of 4 lower than for newborns. For adults, increasing the x-ray tube voltage from 80 to 140 kV improved the CNR for muscle lesions by 130%, for fat lesions by a factor of 2, and for iodine lesions by 25%. As the size of the patient increased from newborn to adults, lesion CNR was reduced by about a factor of 2. The mAs value can be reduced by 80% when scanning newborns while maintaining the same lesion CNR as for adults. Maintaining the CNR of an iodine lesion at a constant level, use of 140 kV increases the energy imparted to an adult patient by nearly a factor of 3.5 in comparison to 80 kV. For fat and muscle lesions, raising the x-ray tube voltage from 80 to 140 kV at a constant CNR increased the patient dose by 37% and 7%, respectively. Our two key findings are that for head CT examinations performed at a constant CNR, the mAs can be substantially reduced when scanning infants, and that use of the lowest x-ray tube voltage will generally reduce patient doses.
我们研究了患者头部特征以及X射线技术因素的选择如何影响计算机断层扫描(CT)图像中的病变对比度和噪声值。从新生儿到成人的五类患者的头部CT图像中获取头部大小和平均亨氏单位(HU)值。使用管电压范围为80至140 kV的X射线光谱来计算平均光子能量以及透过不同大小患者头部的能量注量。确定相对于均匀水背景的脂肪、肌肉和碘病变的图像对比度以及相应的对比度噪声比(CNR)。在保持每个病变的图像CNR恒定的情况下,还计算了给予患者的能量,以确定使辐射剂量最小化的X射线管电压。对于成年人,将管电压从80 kV增加到140 kV时,碘的HU从2.62×10⁵变为1.27×10⁵,脂肪的HU从-1️38变为-108,肌肉的HU从37.1变为33.0。将X射线管电压从80 kV增加到140 kV,能量注量传输百分比增加了高达2倍。对于固定的X射线管电压,成年人的传输能量注量百分比比新生儿低4倍多。对于成年人,将X射线管电压从80 kV增加到140 kV时,肌肉病变的CNR提高了130%,脂肪病变提高了2倍,碘病变提高了25%。随着患者体型从新生儿增加到成年人,病变CNR降低了约2倍。扫描新生儿时,在保持与成人相同病变CNR的情况下,可以将mAs值降低80%。将碘病变的CNR保持在恒定水平,与80 kV相比,使用140 kV时给予成年患者的能量增加了近3.5倍。对于脂肪和肌肉病变,在恒定CNR下将X射线管电压从80 kV提高到140 kV,患者剂量分别增加了37%和7%。我们的两个关键发现是,对于以恒定CNR进行的头部CT检查,扫描婴儿时mAs可以大幅降低,并且使用最低的X射线管电压通常会降低患者剂量。