[一种用于术中心内膜标测的新型球囊电极]

[A new balloon electrode for intraoperative endocardial mapping].

作者信息

Misaki T, Ohtake H, Matsumoto I, Matsunaga Y, Watanabe Y

机构信息

Department of Surgery (1), Kanazawa University School of Medicine.

出版信息

Kyobu Geka. 1992 Aug;45(9):794-6.

PMID:1507706
Abstract

To enhance the utility of epicardial mapping in the intraoperative determination of the location of the arrhythmic source, we developed a new balloon electrode covered with Presnet-tubular dressing material including 46 snap electrode recording sites. The bipolar recording sites, 1.0 mm in diameter and spaced at intervals of 1.5 mm, are constructed of gold and directly attached steel wire. This new balloon electrode was effective intraoperative determination of the lesion in 2 patients one with Wolff-Parkinson-White syndrome and the other with ventricular tachycardia.

摘要

为提高心外膜标测在术中确定心律失常源位置的效用,我们研发了一种新型球囊电极,其覆盖有包含46个扣合电极记录位点的Presnet管状敷料材料。双极记录位点直径为1.0毫米,间隔1.5毫米,由金和直接连接的钢丝构成。这种新型球囊电极在2例患者(1例患有预激综合征,另1例患有室性心动过速)的术中病变确定中有效。