Horn K, Haux P, Kattermann R, Oellerich M, Knoke A, Sommer R, Weissmann A, Vockenhuber M, Clark D, Ng R H
Medizinische Klinik Innenstadt, Universität München, Federal Republic of Germany.
Wien Klin Wochenschr Suppl. 1992;191:80-5.
The present paper describes the multicenter evaluation of the CEDIA Cortisol test for total cortisol. The observed linearity of the test was between 1.2 and 50 micrograms/dL cortisol. The limit of detection was calculated as 1.2/dL. Imprecision studies covering the diagnostically relevant range (5-20 micrograms/dL cortisol) yielded coefficients of variation between 1.7-8.9% (within-run) and 2.7-10.5% (between-day). An interlaboratory survey using 41 human samples and three control sera demonstrated that the new CEDIA Cortisol assay has a good interlaboratory transferability. Method comparison studies between the CEDIA Cortisol test and EIA, FIA, FPIA, and various RIAs yielded an acceptable level of agreement and concordant results in most cases. Low cross-reactivity of the antibody used in the new cortisol assay was observed with precursors or metabolites of cortisol. Especially, dexamethasone did not cross-react. However, prednisolone, 6-methylprednisone, and corticosterone showed cross-reactivities. No limitation by endogenous interferences was observed. The CEDIA Cortisol assay permits the precise, fast and sufficiently specific determination of cortisol. Furthermore, it offers the advantages of a non-radioactive assay and can be performed conveniently on Boehringer Mannheim/Hitachi analyzers in combination with routine clinical chemistry.
本文描述了对CEDIA皮质醇总皮质醇检测法的多中心评估。该检测法观察到的线性范围为1.2至50微克/分升皮质醇。检测限计算为1.2/分升。覆盖诊断相关范围(5至20微克/分升皮质醇)的不精密度研究得出批内变异系数在1.7%至8.9%之间,批间变异系数在2.7%至10.5%之间。一项使用41份人类样本和三份对照血清的实验室间调查表明,新的CEDIA皮质醇检测法具有良好的实验室间转移性。CEDIA皮质醇检测法与酶免疫分析(EIA)、荧光免疫分析(FIA)、荧光偏振免疫分析(FPIA)及各种放射免疫分析(RIA)之间的方法比较研究在大多数情况下得出了可接受的一致性水平和一致结果。新皮质醇检测法中使用的抗体与皮质醇的前体或代谢物的交叉反应性较低。特别是,地塞米松无交叉反应。然而,泼尼松龙、6-甲基泼尼松和皮质酮显示出交叉反应性。未观察到内源性干扰的限制。CEDIA皮质醇检测法可精确、快速且足够特异的测定皮质醇。此外,它具有非放射性检测法的优点,并且可以方便地在Boehringer Mannheim/Hitachi分析仪上结合常规临床化学进行检测。