一种新型视觉模拟量表(VAS)装置:一种基于VAS设计的用于量化主观视觉体验的仪器。

A novel visual analogue scale (VAS) device: an instrument based on the VAS designed to quantify the subjective visual experience.

作者信息

Pointer Jonathan S

机构信息

Optometric Research, 4 Market Square, Higham Ferrers, Northants NN10 8BP, UK.

出版信息

Ophthalmic Physiol Opt. 2004 May;24(3):181-5. doi: 10.1111/j.1475-1313.2004.00180.x.

Abstract

PURPOSE

To evaluate a novel device intended to facilitate the direct quantification of the subjective visual experience. The design principle is based on that of the visual analogue scale (VAS) technique but obviates the need for subsequent measurement unlike the administration of the conventional paper VAS.

METHODS

The visual experience of 134 normally sighted 17-40-year-olds was quantified using (in randomised sequence) a paper VAS and the novel VAS device. A notional 100-point scale was utilised in either case, and the extreme descriptive anchors for both instruments were 'dreadful' and 'perfect'.

RESULTS

The degree of clinical agreement between the two alternative VAS techniques was very high. A small bias (mean = +0.7 VAS units: 95% confidence interval 0.3-1.2 units) towards the paper VAS was evident in these data, an outcome of no clinical significance or impediment as regards the substitution of the novel device for the conventional paper-based approach.

CONCLUSIONS

The moving of a bead along a wire is shown to be a clinically and statistically reliable alternative procedure to the pen-marking of a conventional paper VAS for the evaluation of the recent subjective visual experience, provided that the same minimum/maximum descriptive anchors are used for either instrument. The instant quantification of the VAS score afforded by the novel technique introduced here broadens the potential clinical application of this technique.

摘要

目的

评估一种旨在促进主观视觉体验直接量化的新型设备。其设计原理基于视觉模拟量表(VAS)技术,但与传统纸质VAS的使用不同,该设备无需后续测量。

方法

使用纸质VAS和新型VAS设备(随机顺序)对134名17至40岁视力正常者的视觉体验进行量化。两种情况下均采用100分的概念量表,两种工具的极端描述性锚点均为“糟糕”和“完美”。

结果

两种VAS技术之间的临床一致性程度非常高。这些数据中明显存在向纸质VAS的小偏差(均值=+0.7 VAS单位:95%置信区间0.3 - 1.2单位),就用新型设备替代传统纸质方法而言,这一结果在临床意义或阻碍方面均无影响。

结论

对于评估近期主观视觉体验,沿金属丝移动珠子被证明是一种在临床和统计学上可靠的替代传统纸质VAS笔标记的方法,前提是两种工具使用相同的最小/最大描述性锚点。此处引入的新技术对VAS评分的即时量化拓宽了该技术的潜在临床应用。