Massy-Westropp Nicola, Rankin Wayne, Ahern Michael, Krishnan Jegan, Hearn Trevor C
Department of Orthopaedic Surgery, Daw Park, SA, Australia.
J Hand Surg Am. 2004 May;29(3):514-9. doi: 10.1016/j.jhsa.2004.01.012.
To determine reference ranges for peak, average, and final adult grip strength over 10 seconds by using an electronic dynamometer, and to compare results from hydraulic and electronic dynamometers.
The hand-grip strengths of 476 healthy adult subjects were tested using the electronic (Grippit; AB Detektor, Goteborg, Sweden) and hydraulic (Jamar; Smith and Nephew, Memphis, TN) dynamometers.
Age- and gender-specific reference ranges for the Jamar and Grippit dynamometers are presented. Bland-Altman analysis of the differences between the results obtained using the 2 instruments revealed a bias (mean difference) of 22 N (Jamar - Grippit) and limits of agreement of -86 to 129 N (mean +/- 2 SD), which indicates that grip measurements may vary by up to 215 N between instruments.
The study yielded population reference ranges of peak, average, and final strength over a 10-second grip assessment using an electronic dynamometer. Results from the Grippit and Jamar dynamometers are similar; however, the dynamometers cannot be interchanged. The Grippit provides information about endurance and fatigue of grip over 10 seconds, showing differences between right- and left-dominant adults.
使用电子测力计确定成人10秒内握力峰值、平均握力和最终握力的参考范围,并比较液压测力计和电子测力计的测试结果。
使用电子测力计(Grippit;AB Detektor,瑞典哥德堡)和液压测力计(Jamar;Smith and Nephew,美国田纳西州孟菲斯)对476名健康成人受试者的握力进行测试。
给出了Jamar和Grippit测力计按年龄和性别的参考范围。对两种仪器测试结果差异的Bland-Altman分析显示,偏差(平均差异)为22 N(Jamar - Grippit),一致性界限为-86至129 N(均值±2标准差),这表明不同仪器间的握力测量值可能相差高达215 N。
本研究得出了使用电子测力计进行10秒握力评估时峰值、平均和最终力量的人群参考范围。Grippit和Jamar测力计的结果相似;然而,两种测力计不能互换使用。Grippit可提供10秒内握力耐力和疲劳的信息,显示出右利手和左利手成年人之间的差异。