Viktorova O S, Kogan V T, Manninen S A, Kotiaho T, Ketola R A, Dubenskii B M, Parinov S P, Smirnov O V
Plasma Division, Nuclear and Space Physics Laboratory, A. F. Ioffe Physico-Technical Institute, St. Petersburg, Russia.
J Am Soc Mass Spectrom. 2004 Jun;15(6):823-31. doi: 10.1016/j.jasms.2004.02.002.
Sudden sampling introduction into a membrane inlet mass spectrometer (MIMS) considerably improves the selectivity of the membrane inlet and is therefore applicable even for compounds with low permeabilities through a silicone membrane. In this study the basics of cyclic non-steady-state sudden increase sample injection were studied using a three-membrane inlet and a portable sector double-focusing mass spectrometer. The operational parameters of the inlet system providing the most efficient enrichment of volatile organic compounds (VOCs) in air were defined. Simulation of the diffusion process following sudden sample introduction into the three-membrane inlet was also carried out. Experimental testing of the three-membrane inlet system with the cyclic sudden sample injection mode for benzene, toluene, styrene, and xylene in air was performed. The simulation and the experimental results demonstrated that, when this mode is used, the VOCs/nitrogen relative enrichment factor of samples introduced into the mass spectrometer equipped with a three-membrane inlet is increased by a factor of approximately 10(5) compared with a direct introduction method. This effect may be used to decrease detection limits of compounds obtained with mass spectrometry to decrease matrix flow through the inlet at the same detection limits.
将样品突然引入膜进样质谱仪(MIMS)可显著提高膜进样的选择性,因此即使对于透过硅膜渗透率较低的化合物也适用。在本研究中,使用三膜进样口和便携式扇形双聚焦质谱仪研究了循环非稳态突然增加进样的基本原理。确定了能最有效地富集空气中挥发性有机化合物(VOCs)的进样系统操作参数。还对样品突然引入三膜进样口后的扩散过程进行了模拟。对配备循环突然进样模式的三膜进样口系统进行了空气中苯、甲苯、苯乙烯和二甲苯的实验测试。模拟和实验结果表明,采用这种模式时,与直接进样法相比,引入配备三膜进样口的质谱仪中的样品的VOCs/氮气相对富集因子提高了约10⁵倍。这种效应可用于降低质谱法获得的化合物检测限,以在相同检测限下减少通过进样口的基质流量。