González-Méijome J M, Jorge J, Queiros A, Almeida J B, Parafita M A
Department of Physics (Optometry), School of Sciences, University of Minho, 4710-057 Braga, Portugal.
Ophthalmic Physiol Opt. 2004 Sep;24(5):391-8. doi: 10.1111/j.1475-1313.2004.00203.x.
The purpose of this study was to evaluate if corneal topogometry measurements as taken with the ARK-700A autokeratometer are comparable with those measured by the Medmont E300 videokeratoscope at corresponding locations.
Central and peripheral radius and eccentricity were measured in 122 right eyes of young normal subjects using autokeratometry and videokeratoscopy, obtained in a random order.
Curvature measurements obtained with the ARK700A and Medmont E300 correlated well for the central cornea. Larger differences were observed between peripheral autokeratometry readings and the empirically determined corresponding locations with the videokeratoscope. Correlations between the instruments are inconsistent, resulting in unacceptable confidence intervals. Corneal eccentricity was significantly different between the instruments for the vertical (t = 2.4; p = 0.018) and for the horizontal meridians. In the first case, the difference between the averaged values was not clinically significant, but in the horizontal meridian the AK significantly overestimated eccentricity values (t = -11.5; p < 0.001) with differences which were clinically significant.
Central corneal curvature data obtained by ARK700A and Medmont E300 can be interchanged but the same is not true of peripheral determinations of corneal shape. ARK700A probably measures peripheral corneal shape within an elliptical region between 5 and 7 mm in diameter, with the major axis in the vertical meridian.
本研究旨在评估使用ARK - 700A自动角膜曲率计进行的角膜地形图测量结果与Medmont E300角膜地形图仪在相应位置测量的结果是否具有可比性。
对122名年轻正常受试者的右眼,以随机顺序使用自动角膜曲率计和角膜地形图仪测量中央和周边半径及偏心率。
ARK700A和Medmont E300获得的中央角膜曲率测量值相关性良好。周边自动角膜曲率计读数与角膜地形图仪根据经验确定的相应位置之间存在较大差异。两种仪器之间的相关性不一致,导致置信区间不可接受。两种仪器在垂直子午线(t = 2.4;p = 0.018)和水平子午线的角膜偏心率存在显著差异。在第一种情况下,平均值之间的差异无临床意义,但在水平子午线,自动角膜曲率计显著高估了偏心率值(t = -11.5;p < 0.001),差异具有临床意义。
ARK700A和Medmont E300获得的中央角膜曲率数据可以互换,但周边角膜形状的测定情况并非如此。ARK700A可能测量的是直径在5至7毫米之间、长轴在垂直子午线的椭圆形区域内的周边角膜形状。