Magee T R, Niblett P G, Campbell W B
Department of Surgery, Royal Devon and Exeter Hospital, Wonford, U.K.
Eur J Vasc Surg. 1992 Jan;6(1):21-5. doi: 10.1016/s0950-821x(05)80089-1.
Reinforced vascular grafts are designed to resist compression and flexion forces, for example in axillo-femoral or below-knee femoropopliteal reconstruction. This study compared eight types of 6-mm reinforced grafts (five polytetrafluoroethylene (PTFE) and 3 Dacron). On a specially designed rig, graft diameter and flow were measured during compression by a standard series of weights up to 3.6 kg, and on another rig grafts were flexed through 135 degrees. Meadox Microvel graft resisted compression best (p less than 0.05) but there were no important differences between the other grafts (Vascutek; Bard; Goretex Standard, Thin Walled and Removable Ring; Impra Standard and Thin Wall). Flexion produced no significant changes in flow through the grafts. In addition, five general surgeons were asked to score the ease of removal of reinforcement, passage of sutures, and the tendency of sutures to cut out of each graft. Removal of reinforcement and passage of sutures were easier with the PTFE grafts, but the Dacron was more resistant to sutures cutting out. This study shows that these grafts resist flexion and compression well. Their different handling characteristics and data on patency at different sites should remain the main determinants in choice of graft.
强化血管移植物旨在抵抗压缩和弯曲力,例如在腋-股或膝下股-腘重建中。本研究比较了8种6毫米的强化移植物(5种聚四氟乙烯(PTFE)和3种涤纶)。在一个专门设计的装置上,通过一系列标准重量直至3.6千克的重物对移植物进行压缩时,测量其直径和流量;在另一个装置上,将移植物弯曲135度。Meadox Microvel移植物抵抗压缩的能力最强(p小于0.05),但其他移植物(Vascutek;Bard;Goretex标准型、薄壁型和可移除环型;Impra标准型和薄壁型)之间没有显著差异。弯曲对移植物内的血流没有产生显著变化。此外,邀请了5位普通外科医生对每种移植物的强化物移除难易程度、缝线穿过情况以及缝线切割穿出的倾向进行评分。PTFE移植物的强化物移除和缝线穿过更容易,但涤纶移植物对缝线切割穿出的抵抗力更强。本研究表明,这些移植物能很好地抵抗弯曲和压缩。它们不同的操作特性以及不同部位通畅性的数据应仍是选择移植物的主要决定因素。