Abstract

Over the past several years, digital mammography systems have been installed clinically across North America in small but growing numbers. A photostimulable phosphor-based full-field digital mammography image was evaluated in this investigation. Commonly known as computed radiography (CR), its use closely mimics the screen-film mammography paradigm. System performance using modulation transfer function (MTF) and detective quantum efficiency (DQE) metrics show MTF(2.5 mm(-1)) = 0.5, DQE(2.5 mm(-1)) = 0.3, and MTF(5.0 mm(-1)) = 0.2, DQE(5.0 mm(-1)) = 0.05, for a 26 kVp beam, 0.03 mm molybdenum tube filtration, 4.5 cm tissue attenuation, and 15 mR incident exposure to the detector. Slightly higher DQE values were measured at 32 kVp with 0.025 mm rhodium tube filtration. CR mammography advantages include the ability to use existing mammography machines, where multiple rooms can be converted to "digital" operation, which allows overall cost savings compared to integrated digital mammography systems. Chief disadvantages include the labor-intensive handling of the cassettes prior to and after the imaging exam, lack of a direct interface to the x-ray system for recording technique parameters, and relatively slow processing time. Clinical experience in an IRB-approved research trial has suggested that digital mammography with photostimulable storage phosphors and a dedicated CR reader is a viable alternative to conventional screen-film mammography.

摘要

在过去几年中,数字乳腺摄影系统已在北美临床上少量但不断增加地安装。在本研究中对基于光激励荧光体的全场数字乳腺摄影图像进行了评估。它通常被称为计算机X线摄影(CR),其使用方式与屏-片乳腺摄影模式非常相似。使用调制传递函数(MTF)和探测量子效率(DQE)指标的系统性能表明,对于26 kVp的射线束、0.03 mm钼靶管滤过、4.5 cm组织衰减以及15 mR的探测器入射曝光量,MTF(2.5 mm(-1)) = 0.5,DQE(2.5 mm(-1)) = 0.3,MTF(5.0 mm(-1)) = 0.2,DQE(5.0 mm(-1)) = 0.05。在32 kVp、0.025 mm铑靶管滤过的情况下测量到的DQE值略高。CR乳腺摄影的优点包括能够使用现有的乳腺摄影设备,多个房间可以转换为“数字”操作,与集成数字乳腺摄影系统相比可节省总体成本。主要缺点包括成像检查前后暗盒处理劳动强度大、缺乏与X射线系统记录技术参数的直接接口以及处理时间相对较慢。一项经机构审查委员会批准的研究试验中的临床经验表明,使用光激励存储荧光体和专用CR读取器的数字乳腺摄影是传统屏-片乳腺摄影的可行替代方案。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验