Effects of x-ray spectra on the DQE of a computed radiography system.

作者信息

Fetterly K A, Hangiandreou N J

机构信息

Diagnostic Radiology, Mayo Clinic, Rochester, Minnesota 55905, USA.

出版信息

Med Phys. 2001 Feb;28(2):241-9. doi: 10.1118/1.1339883.

Abstract

The effect of incident x-ray beam quality on the measured detective quantum efficiency (DQE) of a computed radiography system was investigated. The incident x-ray beams used had peak tube potentials of 70, 95, and 120 kVp, were filtered with various thicknesses of a "patient equivalent phantom" (PEP), aluminum, and copper, and provided a consistent exposure to the storage phosphor. For each peak tube potential and filter combination, the one-dimensional modulation transfer function and noise power spectrum were measured and the square of the incident signal-to-noise ratio was estimated. The spatial frequency dependent DQE was calculated from these data. The DQE was integrated to provide an overall estimate of the efficiency and frequency response of the computed radiography system for the various x-ray beams. There was found to be a wide range of integral DQE (IDQE) values for the peak tube potential and filter combinations used. For example, the IDQE ranged from 3.0 to 0.9 mm(-2) using the peak tube potential and filter combinations 70 kVp with 5.1 cm PEP and 120 kVp with 30.3 cm PEP, respectively. Finally, peak tube potential and filter combinations 70 kVp with 10.2 cm PEP and 120 kVp with 20.2 cm PEP were chosen as standard x-ray beams that will be used at our facility to measure the DQE of digital radiographic imaging systems for evaluation and acceptance testing.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验