Influence of x-ray pulse parameters on the image quality for moving objects in digital cardiac imaging.

作者信息

Guibelalde Eduardo, Vano Eliseo, Vaquero Francisco, González Luciano

机构信息

Medical Physics Group, Department of Radiology, University Complutense of Madrid, 28040 Madrid, Spain.

出版信息

Med Phys. 2004 Oct;31(10):2819-25. doi: 10.1118/1.1790091.

Abstract

The image quality of a single frame in a modern cardiac imaging x-ray facility can be improved by adjusting the automatic pulse exposure parameters. The effects of acquisition rate on patient dose and the detectability of moving objects have been fully described in scientific literature. However, the influence of automatic pulse exposure parameters is still to be determined. Images of a moving wheel (with lead wires) were acquired using an H5000 Philips Integris cardiac x-ray system. Poly(methylmethacrylate) plastic samples 20 and 30 cm thick were employed as the build-up phantom to simulate a patient. The images were obtained using preset clinical parameters for cardiac imaging procedures. The signal detectability and motion blur of a contrast bar at a transversal speed in the range of 100-150 mm/s were evaluated with a cine pulse width of 3, 5, 7, and 10 ms under automatic mA kV regulation. Two levels of exposure at the image intensifier entrance were included in this study. Signal detectability was analyzed in terms of the signal-to-noise ratio (SNR) and the value of SNR2/entrance surface dose. The blurring was modeled as a Gaussian-shaped blurring function, and the motion blur was expressed in terms of the peak full width at half maximum and amplitude (apparent contrast) of the resolution functions. A contrast bar simulating a vessel in motion at the maximum velocities of typical cardiac structures was exposed. Severe loss of image quality occurred at pulse widths > or =7 ms. It is also shown that below 5 ms static nonlinearities, likely caused by the need to use a large focus for cine acquisition, dominate the blurring process.

摘要

通过调整自动脉冲曝光参数,可以提高现代心脏成像X射线设备中单个帧的图像质量。采集速率对患者剂量和移动物体可检测性的影响已在科学文献中得到充分描述。然而,自动脉冲曝光参数的影响仍有待确定。使用飞利浦Integris H5000心脏X射线系统采集了一个带引线的转动轮的图像。采用20厘米和30厘米厚的聚甲基丙烯酸甲酯塑料样品作为等效体模来模拟患者。使用心脏成像程序的预设临床参数获取图像。在自动毫安千伏调节下,在3、5、7和10毫秒的电影脉冲宽度下,评估了横向速度在100 - 150毫米/秒范围内的对比条的信号可检测性和运动模糊。本研究包括图像增强器入口处的两个曝光水平。根据信噪比(SNR)和SNR2/入口表面剂量值分析信号可检测性。模糊被建模为高斯形状的模糊函数,运动模糊用分辨率函数的半高宽峰值和幅度(表观对比度)表示。曝光了一个模拟典型心脏结构最大速度下运动血管的对比条。在脉冲宽度大于或等于7毫秒时,图像质量出现严重下降。研究还表明,在5毫秒以下,可能由于电影采集需要使用大焦点而导致的静态非线性主导了模糊过程。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验