LXi Research, Division of Medical Physics, University of Leeds, Worsley Building, Clarendon Way, Leeds LS2 9JT, United Kingdom.
Med Phys. 2010 Oct;37(10):5258-69. doi: 10.1118/1.3488911.
The aim of this research was to explore x-ray beam parameters with intent to optimize pediatric x-ray settings in the cardiac catheterization laboratory. This study examined the effects of peak x-ray tube voltage (kVp) and of copper (Cu) x-ray beam filtration independently on the image quality to dose balance for pediatric patient sizes. The impact of antiscatter grid removal on the image quality to dose balance was also investigated.
Image sequences of polymethyl methacrylate phantoms approximating chest sizes typical of pediatric patients were captured using a modern flat-panel receptor based x-ray imaging system. Tin was used to simulate iodine-based contrast medium used in clinical procedures. Measurements of tin detail contrast and flat field image noise provided the contrast to noise ratio. Entrance surface dose (ESD) and effective dose (E) measurements were obtained to calculate the figure of merit (FOM), CNR2/dose, which evaluated the dose efficiency of the x-ray parameters investigated. The kVp, tube current (mA), and pulse duration were set manually by overriding the system's automatic dose control mechanisms. Images were captured with 0, 0.1, 0.25, 0.4, and 0.9 mm added Cu filtration, for 50, 55, 60, 65, and 70 kVp with the antiscatter grid in place, and then with it removed.
For a given phantom thickness, as the Cu filter thickness was increased, lower kVp was favored. Examining kVp alone, lower values were generally favored, more so for thinner phantoms. Considering ESD, the 8.5 cm phantom had the highest FOM at 50 kVp using 0.4 mm of Cu filtration. The 12 cm phantom had the highest FOM at 55 kVp using 0.9 mm Cu, and the 16 cm phantom had highest FOM at 55 kVp using 0.4 mm Cu. With regard to E, the 8.5 and 12 cm phantoms had the highest FOM at 50 kVp using 0.4 mm of Cu filtration, and the 16 cm phantom had the highest FOM at 50 kVp using 0.25 mm Cu. Antiscatter grid removal improved the FOM for a given set of x-ray conditions. Under aforesaid optimal settings, the 8.5 cm phantom FOM improved by 24% and 33% for ESD and E, respectively. Corresponding improvements were 26% and 24% for the 12 cm phantom and 6% and 15% for the 16 cm phantom.
For pediatric patients, using 0.25-0.9 mm Cu filtration in the x-ray beam while maintaining 50-55 kVp, depending on patient size, provided optimal x-ray image quality to dose ratios. These settings, adjusted for x-ray tube loading limits and clinically acceptable image quality, should provide a useful strategy for optimizing iodine contrast agent based cardiac x-ray imaging. Removing the antiscatter grid improved the FOM for the 8.5 and 12 cm phantoms, therefore grid removal is recommended for younger children. Improvement for the 16 cm phantom declined into the estimated margin of error for the FOM; the need for grid removal for older children would depend on practical feasibility in the clinical environment.
本研究旨在探讨 X 射线束参数,以优化心导管室中的儿科 X 射线设置。本研究分别检查了峰值 X 射线管电压(kVp)和铜(Cu)X 射线束过滤对小儿患者尺寸的图像质量与剂量平衡的影响。还研究了去除防散射栅格对图像质量与剂量平衡的影响。
使用现代平板探测器基于 X 射线成像系统捕获模拟儿科患者典型胸部大小的聚甲基丙烯酸甲酯(PMMA)体模的图像序列。使用锡模拟临床操作中使用的碘基对比剂。锡细节对比度和平面场图像噪声的测量提供了对比噪声比。获取入射表面剂量(ESD)和有效剂量(E)测量值,以计算评估所研究 X 射线参数的效率的性能指标(FOM),即对比噪声比 2/剂量(CNR2/dose)。kVp、管电流(mA)和脉冲持续时间通过手动覆盖系统的自动剂量控制机制进行设置。在有防散射栅格和去除防散射栅格的情况下,对 50、55、60、65 和 70 kVp 分别使用 0、0.1、0.25、0.4 和 0.9 mm 附加 Cu 过滤进行了图像捕获。
对于给定的体模厚度,随着 Cu 滤光片厚度的增加,较低的 kVp 更受欢迎。单独检查 kVp 时,通常更倾向于较低的值,对于较薄的体模更是如此。考虑到 ESD,在使用 0.4mm Cu 过滤的情况下,8.5cm 体模在 50kVp 时具有最高的 FOM。在使用 0.9mm Cu 的情况下,12cm 体模在 55kVp 时具有最高的 FOM,在使用 0.4mm Cu 的情况下,16cm 体模在 55kVp 时具有最高的 FOM。关于 E,8.5 和 12cm 体模在使用 0.4mm Cu 过滤的情况下在 50kVp 时具有最高的 FOM,而 16cm 体模在使用 0.25mm Cu 的情况下在 50kVp 时具有最高的 FOM。去除防散射栅格改善了给定 X 射线条件的 FOM。在上述最佳设置下,8.5cm 体模的 ESD 和 E 的 FOM 分别提高了 24%和 33%。对于 12cm 体模,相应的改进分别为 26%和 24%,对于 16cm 体模,分别为 6%和 15%。
对于儿科患者,根据患者的体型,在 X 射线束中使用 0.25-0.9mm Cu 过滤,同时保持 50-55kVp,可以提供最佳的 X 射线图像质量与剂量比。这些根据 X 射线管负载限制和临床可接受的图像质量进行调整的设置,应为优化碘对比剂为基础的心脏 X 射线成像提供了一种有用的策略。去除防散射栅格改善了 8.5 和 12cm 体模的 FOM,因此建议对年龄较小的儿童去除栅格。对于 16cm 体模,改善情况下降到 FOM 的估计误差范围内;对于年龄较大的儿童,是否需要去除栅格将取决于临床环境中的实际可行性。