Shastri K A, Logue G L
Department of Medicine, State University of New York, Buffalo.
ASAIO J. 1992 Jan-Mar;38(1):47-51. doi: 10.1097/00002480-199201000-00011.
Quantitation of complement activation by polyacrilonitrile (PAN) dialyzer membrane is complicated by the high adsorptive capacity of the membrane for fluid phase anaphylotoxins. Assays for these anaphylotoxins, therefore, underestimate the degree of complement activation produced by this membrane. Alternative methods of measuring in vitro complement activation by the PAN and Cuprophan membranes were explored by incubating normal human erythrocytes with the membranes in the presence of serum. This led to deposition of C3d on these "innocent bystander" red cells, and provided an independent parameter for measuring complement activation. The PAN membrane caused significantly more C3d deposition on red cells, and thus more complement activation than Cuprophan. The possible significance of complement activation by PAN membrane, in consideration of its property of binding the resultant anaphylotoxins, is discussed.
聚丙烯腈(PAN)透析器膜对补体激活的定量分析因该膜对液相过敏毒素的高吸附能力而变得复杂。因此,针对这些过敏毒素的检测低估了该膜产生的补体激活程度。通过在血清存在的情况下将正常人红细胞与膜一起孵育,探索了测量PAN膜和铜仿膜体外补体激活的替代方法。这导致C3d沉积在这些“无辜旁观者”红细胞上,并提供了一个测量补体激活的独立参数。与铜仿膜相比,PAN膜导致红细胞上的C3d沉积明显更多,因此补体激活也更多。考虑到PAN膜结合所产生的过敏毒素的特性,讨论了其补体激活的可能意义。