[Image characteristics of the computer tomograph (author's transl)].

作者信息

Schultz E, Felix R, Lackner K, Bergeder H D, Thurn P

出版信息

Rofo. 1979 Apr;130(4):479-86. doi: 10.1055/s-0029-1231313.

Abstract

In the present paper the images produced by spheres of varying diameter (d = 4,6,8,10 mm) embedded in a homogeneous substance of varying densities (H' = 3,48,93,137 Hounsfield units) as produced by computer tomography were studied. Measurements show that for each sphere diameter there is a minimum contrast between the substance of the sphere and its surroundings which must exist in order to show this sphere by computer tomography. It was also shown that the partial volume effect is of significance for the attenuation of the spheres. The effect of the position of the spheres within the slices to be imaged was studied. The results clearly indicate that in the investigation of small tissue changes in the body, the thickness of the slice and patient displacement should be kept as small as possible.