4特斯拉下乳腺1H磁共振波谱中呼吸诱导的B0变化的测量与校正

Measurement and correction of respiration-induced B0 variations in breast 1H MRS at 4 Tesla.

作者信息

Bolan Patrick J, Henry Pierre-Gilles, Baker Eva H, Meisamy Sina, Garwood Michael

机构信息

Center for Magnetic Resonance Research, University of Minnesota School of Medicine, Minneapolis 55455, USA.

出版信息

Magn Reson Med. 2004 Dec;52(6):1239-45. doi: 10.1002/mrm.20277.

Abstract

Respiratory motion is well known to cause artifacts in magnetic resonance spectroscopy (MRS). In MRS of the breast, the dominant artifact is not due to motion of the breast itself, but rather it is produced by B0 field distortions associated with respiratory motion of tissues in the chest and abdomen. This susceptibility artifact has been reported to occur in the brain, but it is more apparent in the breast due to the anatomic proximity of the lungs. In the breast, these B0 distortions cause shot-to-shot frequency shifts, which vary an average of 24 Hz during a typical 1H MRS scan at 4 T. This variation can be corrected retrospectively by frequency shifting individual spectra prior to averaging. If not corrected, these shifts reduce spectral resolution and increase peak fitting errors. This work demonstrates the artifact, describes a method for correcting it, and evaluates its impact on quantitative spectroscopy. When the artifact is not corrected, quantification errors increase by an average of 28%, which dramatically impacts the ability to measure metabolite resonances at low signal-to-noise ratios.

摘要

众所周知,呼吸运动会在磁共振波谱(MRS)中产生伪影。在乳腺的MRS中,主要伪影并非由乳腺自身的运动引起,而是由与胸部和腹部组织呼吸运动相关的B0场畸变产生的。这种敏感性伪影在脑部也有报道,但由于肺部与乳腺在解剖位置上接近,所以在乳腺中更为明显。在乳腺中,这些B0畸变会导致逐次扫描的频率偏移,在4T场强下进行典型的1H MRS扫描时,平均频率偏移为24Hz。这种变化可以通过在平均之前对各个谱进行频率偏移来进行回顾性校正。如果不进行校正,这些偏移会降低谱分辨率并增加峰拟合误差。这项工作展示了这种伪影,描述了一种校正方法,并评估了其对定量波谱的影响。当不校正伪影时,定量误差平均增加28%,这对在低信噪比下测量代谢物共振的能力有显著影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验