使用静脉报警控制(IVAC)泵进行准确的骨筋膜室压力测量。一种技术报告。

Accurate compartment pressure measurement using the Intervenous Alarm Control (IVAC) Pump. Report of a technique.

作者信息

Uppal G S, Smith R C, Sherk H H, Mooar P

机构信息

Department of Orthopaedic Surgery, Medical College of Pennsylvania, Philadelphia 19129.

出版信息

J Orthop Trauma. 1992;6(1):87-9.

PMID:1556629
Abstract

This study demonstrates the accuracy of a common, readily available pressure transducer system (IVAC pump) for the measurement of compartment syndromes. An in vitro assay was used to compare the accuracy of the IVAC pump manometer with the Hewlett-Packard monometer (HP78532B) and a miniaturized digital fluid pressure manometer (Stryker). The IVAC pump was accurate within 2.4 mm Hg difference of the HP78532B and Stryker manometers. Twenty-five patients suspected of having compartment syndrome posttibial fracture had compartment pressures measured using the IVAC pump. In all the patients, compartment syndrome was accurately assessed (positive in one case and negative in the remainder). The IVAC pump is readily available in most hospitals. It is easy to use and provides accurate measurements of compartment pressures.

摘要

本研究证明了一种常见的、易于获得的压力传感器系统(IVAC泵)用于测量骨筋膜室综合征的准确性。采用体外试验比较IVAC泵压力计与惠普压力计(HP78532B)和小型数字流体压力计(史赛克)的准确性。IVAC泵与HP78532B和史赛克压力计的差值在2.4毫米汞柱以内时测量准确。25例怀疑胫骨骨折后发生骨筋膜室综合征的患者使用IVAC泵测量骨筋膜室内压力。在所有患者中,均准确评估了骨筋膜室综合征(1例为阳性,其余为阴性)。IVAC泵在大多数医院都很容易获得。它易于使用,能准确测量骨筋膜室内压力。