Comparing the methods of measuring compartment pressures in acute compartment syndrome.

作者信息

Uliasz Annemarie, Ishida Jay T, Fleming Jason K, Yamamoto Loren G

机构信息

Department of Pediatrics, University of Hawaii John A. Burns School of Medicine, Kapiolani Medical Center for Women and Children, Honolulu, Hawaii 96826, USA.

出版信息

Am J Emerg Med. 2003 Mar;21(2):143-5. doi: 10.1053/ajem.2003.50035.

Abstract

The objectives of this study were to compare the reliability and accuracy of published methods of measuring intramuscular compartment pressures. A muscle compartment model was constructed. Pressure was standardized using a water column exerting pressure on a muscle slab. The following three methods of intracompartmental pressure were used to measure the pressure within the muscle slab of the model: (1) Stryker, (2) manometric IV pump, and (3) Whitesides method. The Stryker and IV pump methods were similarly accurate compared with the "standard pressure." The Whitesides method was not reliable in measuring a reproducible pressure. The IV pump method is comparable to the Stryker instrument in terms of accuracy and ease of use. If a Stryker instrument is unavailable, the IV pump method can serve as a suitable and satisfactory alternative.

摘要

本研究的目的是比较已发表的测量肌内室压力方法的可靠性和准确性。构建了一个肌室模型。使用水柱对肌肉平板施加压力来标准化压力。采用以下三种肌室内压力测量方法来测量模型肌肉平板内的压力:(1) 史赛克(Stryker)法,(2) 压力计静脉输液泵法,以及(3) 怀特赛兹(Whitesides)法。与“标准压力”相比,史赛克法和静脉输液泵法的准确性相似。怀特赛兹法在测量可重复的压力方面不可靠。静脉输液泵法在准确性和易用性方面与史赛克仪器相当。如果没有史赛克仪器,静脉输液泵法可作为一种合适且令人满意的替代方法。