[单光子吸收法(¹²⁵I)测量骨矿物质含量的准确性和可重复性。源强度的影响]

[The accuracy and reproducibility of the measurement of bone mineral content by single-photon absorptiometry (125I). The effect of source strength].

作者信息

Cattaneo G M, Rinaldi G P, Rubinacci A

机构信息

Clinica Ortopedica, Istituto Scientifico Ospedale S. Raffaele, Milano.

出版信息

Radiol Med. 1992 Jan-Feb;83(1-2):54-8.

PMID:1557545
Abstract

This study was aimed at verifying whether accuracy and reproducibility of mineral content measurements of bone phantoms by single-photon absorptiometry change as a function of source strength, and whether these changes depend on the phantom's width. To these purposes, the Norland 2780 bone mineralometer was used. With this unit, the mineral contents of two cylindrical phantoms of different diameter and of known mineral content were measured every week, 5 times on the whole, up to exhaustion of the source. Both accuracy and reproducibility decreased as source strength decreased. The lowest reproducibility was observed for the measurement of the smaller phantom with low source strength. It was therefore concluded that when source strength reaches the threshold value of 2000 cps, the measurement is so much affected as to jeopardize the reliability of short-term longitudinal studies of bone mineral content in patients.

摘要

本研究旨在验证单光子吸收法测量骨模型矿物质含量的准确性和可重复性是否随源强度而变化,以及这些变化是否取决于模型的宽度。为此,使用了Norland 2780骨密度仪。用该仪器每周测量两个直径不同、矿物质含量已知的圆柱形模型的矿物质含量,共测量5次,直至源耗尽。随着源强度降低,准确性和可重复性均下降。对于低源强度下较小模型的测量,观察到最低的可重复性。因此得出结论,当源强度达到2000 cps的阈值时,测量受到的影响极大,以至于危及患者骨矿物质含量短期纵向研究的可靠性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验