直线加速器输出的可变性。

Linear accelerator output variability.

作者信息

Luketina I A, Greig L

机构信息

Department of Radiation Therapy, Wellington School of Medicine and Health Sciences, University of Otago, Wellington, New Zealand.

出版信息

Australas Phys Eng Sci Med. 2004 Sep;27(3):155-9. doi: 10.1007/BF03178676.

Abstract

The Wellington Cancer Centre is equipped with two matched linear accelerators (Varian 2100CD) linear accelerators (identified as SN1027 and SN42). Each morning, before treatments commence, a radiation therapist carries out an output constancy check of the radiation output and every fortnight a physicist measures, in a phantom, the delivered radiation dose to check on the machine calibration. The daily output checks have been recorded into a database (Argus QA for Radiation Oncology) since August 1997 and in June 1998 the fortnightly calibration measurements were added. The information in the database, up to April 2003, has been analysed to consider the quality of the daily constancy checks as compared with the fortnightly calibration measurements and whether the data contains useful information on machine performance. After allowance for the effects of machine recalibration the fortnightly calibration measurements had an average standard deviation of 0.4% and the daily constancy checks 0.8%. The daily constancy checks had a greater number of large deviations than would be expected assuming a normal distribution and were not a good predictor of the need for a recalibration. The fortnightly calibration measurements with a much lower spread give a reliable indication of the need for a recalibration allowing the adoption of a +/- 1% tolerance. Over the period analysed one accelerator (SN42) was relatively stable with the output generally drifting between +/- 1% while the other (SN1027) had a consistent increase in the average output of about 2.5% per year.

摘要

惠灵顿癌症中心配备了两台匹配的直线加速器(瓦里安2100CD)(序列号分别为SN1027和SN42)。每天早晨,在治疗开始前,放射治疗师会对辐射输出进行输出稳定性检查,并且每两周有一位物理学家在模体中测量 delivered 辐射剂量以检查机器校准情况。自1997年8月起,每日输出检查记录已录入数据库(放射肿瘤学的阿格斯质量保证系统),1998年6月又增加了每两周一次的校准测量记录。对截至2003年4月数据库中的信息进行了分析,以比较每日稳定性检查与每两周校准测量的质量,以及数据是否包含有关机器性能的有用信息。在考虑机器重新校准的影响后,每两周校准测量的平均标准偏差为0.4%,每日稳定性检查为0.8%。每日稳定性检查出现的大偏差数量比假设为正态分布时预期的要多,并且不是重新校准需求的良好预测指标。每两周校准测量的离散度低得多,能可靠地表明重新校准的需求,从而可采用±1%的公差范围。在所分析的时间段内,一台加速器(SN42)相对稳定,输出一般在±1%之间波动,而另一台(SN1027)平均输出每年持续增加约2.5%。 (注:原文中“delivered radiation dose”这里“delivered”翻译为“ delivered ”存疑,可能有误,可进一步确认准确含义后调整)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验