Anderson A F, Snyder R B, Federspiel C F, Lipscomb A B
Lipscomb Clinic, Nashville, TN 37205.
Am J Sports Med. 1992 Mar-Apr;20(2):135-40. doi: 10.1177/036354659202000207.
We performed a prospective study on 50 subjects with normal knees and 50 patients with chronic unilateral disruption of the anterior cruciate ligament. In a randomized testing sequence, both groups were examined with five arthrometers: the MEDmetric KT-1000, the Stryker Knee Laxity Tester, the Acufex Knee Signature System, the Dyonics Dynamic Cruciate Tester, and the Genucom Knee Analysis System. Each examination was performed according to protocol with the knee at 30 degrees of flexion. The total anterior laxity measurements of the normal subjects using the Dyonics Dynamic Cruciate Tester and Acufex Knee Signature System were approximately half of the KT-1000, Stryker, and Genucom values. A comparison of the side-to-side measurements revealed no statistically significant difference in the values of the five arthrometers. However, the Genucom showed an unacceptably high number of normal subjects with laxity values that suggested an anterior cruciate ligament tear. Total anterior laxity measurements of the anterior cruciate ligament deficient knees were almost twice those of normal knees with each device except the Genucom. Comparison of the mean side-to-side difference in the patients with ACL deficiency demonstrated statistically significant differences in the values recorded with the five arthrometers. The side-to-side difference was greatest with the KT-1000 and least for the KSS. In the 89 N Lachman test, the KT-1000 and Stryker demonstrated the highest diagnostic accuracy. The maximum manual test improved the accuracy of each device. This study establishes the total anterior laxity measurements cannot be generalized from one device to another in either group of subjects.(ABSTRACT TRUNCATED AT 250 WORDS)
我们对50名膝关节正常的受试者和50名慢性单侧前交叉韧带断裂的患者进行了一项前瞻性研究。在随机测试序列中,两组均使用五种关节测量仪进行检查:MEDmetric KT - 1000、史赛克膝关节松弛度测试仪、Acufex膝关节特征系统、戴奥尼克斯动态交叉韧带测试仪和Genucom膝关节分析系统。每次检查均按照方案在膝关节屈曲30度时进行。使用戴奥尼克斯动态交叉韧带测试仪和Acufex膝关节特征系统对正常受试者进行的总前向松弛度测量约为KT - 1000、史赛克和Genucom测量值的一半。两侧测量值的比较显示,五种关节测量仪的值在统计学上无显著差异。然而,Genucom显示出数量多得不可接受的正常受试者,其松弛度值提示前交叉韧带撕裂。除Genucom外,每种设备对前交叉韧带损伤膝关节的总前向松弛度测量几乎是正常膝关节的两倍。对ACL缺损患者平均两侧差异的比较表明,五种关节测量仪记录的值在统计学上有显著差异。KT - 1000的两侧差异最大,KSS的两侧差异最小。在89 N拉赫曼试验中,KT - 1000和史赛克显示出最高的诊断准确性。最大手动试验提高了每种设备的准确性。本研究表明,在两组受试者中,不能将一种设备的总前向松弛度测量结果推广到另一种设备。(摘要截取自250字)