Neurostimulation systems: assessment of magnetic field interactions associated with 1.5- and 3-Tesla MR systems.

作者信息

Baker Kenneth B, Nyenhuis John A, Hrdlicka Greg, Rezai Ali R, Tkach Jean A, Shellock Frank G

机构信息

Department of Neurology, The Cleveland Clinic Foundation, Cleveland, Ohio, USA.

出版信息

J Magn Reson Imaging. 2005 Jan;21(1):72-7. doi: 10.1002/jmri.20224.

Abstract

PURPOSE

To evaluate magnetic field interactions at 1.5- and 3-Tesla for implantable pulse generators (IPGs) and radiofrequency (RF) receivers used for implantable neurostimulation systems.

MATERIALS AND METHODS

Measurements of magnetically induced displacement force and torque were determined for 10 devices (seven IPGs, three RF receivers) used for neurostimulation systems. Displacement force and torque were assessed at various positions in 1.5- and 3-Tesla MR systems using standardized techniques.

RESULTS

Four IPGs exhibited force ratios (magnetic attraction force/device weight) greater than 1.0, with the overall magnitude of the force ratio increasing significantly when comparing the 1.5-Tesla to the 3-Tesla MR system. Of the seven IPGs tested, one exhibited a torque ratio (magnetic induced torque/product of the device weight and length) greater than 1.0. The RF receivers displayed relatively strong magnetic field interactions at both 1.5- and 3-Tesla, exhibiting force and torque ratios greater than 1.0.

CONCLUSIONS

The neurostimulation implants tested exhibited varying degrees of magnetic field interactions, with four of the seven IPGs and the three RF receivers exhibiting at least one MR-induced force or torque value greater than the effect of gravity. These findings have important implications for patients with these implants who are referred for MRI examinations.

摘要

目的

评估用于植入式神经刺激系统的植入式脉冲发生器(IPG)和射频(RF)接收器在1.5特斯拉和3特斯拉磁场中的相互作用。

材料与方法

对用于神经刺激系统的10种设备(7个IPG、3个RF接收器)进行磁致位移力和扭矩测量。使用标准化技术在1.5特斯拉和3特斯拉磁共振系统的不同位置评估位移力和扭矩。

结果

4个IPG的力比(磁吸引力/设备重量)大于1.0,将1.5特斯拉磁共振系统与3特斯拉磁共振系统进行比较时,力比的总体大小显著增加。在测试的7个IPG中,有1个的扭矩比(磁致扭矩/设备重量与长度的乘积)大于1.0。RF接收器在1.5特斯拉和3特斯拉时均表现出相对较强的磁场相互作用,力比和扭矩比均大于1.0。

结论

测试的神经刺激植入物表现出不同程度的磁场相互作用,7个IPG中的4个以及3个RF接收器表现出至少一个磁共振诱导的力或扭矩值大于重力的影响。这些发现对接受MRI检查的植入这些设备的患者具有重要意义。