Suppr超能文献

Assessment of a point-of-care test for paracetamol and salicylate in blood.

作者信息

Dale C, Aulaqi A A M, Baker J, Hobbs R C, Tan M E L, Tovey C, Walker I A L, Henry J A

机构信息

Academic Department of Accident and Emergency Medicine, St Mary's Hospital, London W2 1NY, UK.

出版信息

QJM. 2005 Feb;98(2):113-8. doi: 10.1093/qjmed/hci016. Epub 2005 Jan 17.

Abstract

BACKGROUND

Paracetamol and salicylate are commonly taken in acute overdose. Clinicians have a low threshold for excluding the presence of these two drugs, by ordering laboratory tests in any patient suspected of ingesting an overdose or with an altered mental state.

AIM

To test the effectiveness of a new point of care test that qualitatively detects paracetamol and salicylate in blood and to examine the potential time saved by its use.

DESIGN

Prospective multicentre trial.

METHODS

The new test was compared with laboratory analysis in a routine blood sample taken from patients presenting to emergency departments with suspected overdose.

RESULTS

The test had sensitivities of 98.5% and 88.5%, and specificities of 74.7% and 92%, for paracetamol and salicylate, respectively, at cut-off levels of 25 mg/l and 100 mg/l, respectively The point of care test results were available 2 h before the laboratory result.

DISCUSSION

This point-of-care test could be used to rule out an overdose with either of these two drugs, and could thus lead to earlier clinical decisions for suspected overdose patients. Recommendations have been made following this trial that the cut-off value for paracetamol should be reduced from 25 mg/l to 12.5 mg/l in order to increase its usefulness. To prevent the test being misread, we also suggest that each device should be embossed to remind users that the presence of a line indicates there is no drug present.