Weisskoff R M, Kiihne S
Massachusetts General Hospital, Department of Radiology, Charlestown 02129.
Magn Reson Med. 1992 Apr;24(2):375-83. doi: 10.1002/mrm.1910240219.
We present a novel NMR imaging technique that allows absolute determination of the magnetic susceptibility constant, chi, of a solution. By comparing the phase difference of MR images produced with an instant (echo planar) "offset" spin-echo sequence, we obtain a direct measure of the magnetic field perturbations caused by the solution. We demonstrate this method with Gd(DTPA), Dy(DTPA), human red blood cells, and superparamagnetic iron oxide particles.