We report our observations on MR imaging quality and functionality of a recently introduced intracranial pressure monitoring device. The device was tested at two different field strengths in a pig brain specimen to investigate MR imaging artifacts, probe function during and after MR data acquisition, and device-related temperature changes in the brain tissue. Image reading was not impaired, and probe function, although reduced, was not fully interrupted during data acquisition.
我们报告了对一种最近推出的颅内压监测设备的磁共振成像(MR)质量和功能的观察结果。该设备在猪脑标本中于两种不同场强下进行了测试,以研究MR成像伪影、MR数据采集期间及之后的探头功能,以及脑组织中与设备相关的温度变化。图像读取未受影响,并且探头功能在数据采集期间虽有所降低,但并未完全中断。