Department of Communication Disorders, Penn State University, University Park 16802.
J Am Acad Audiol. 1992 Jan;3(1):51-9.
The purpose of this study was to obtain normal hearing thresholds for an ER-3A insert earphone as a contribution for future standardization and for comparison with previous studies and interim ANSI and ISO insert earphone reference equivalent threshold sound pressure levels (RETSPLs). Hearing thresholds were obtained on each ear of 48 normal-hearing subjects from 125 to 8000 Hz using ER-3A and TDH-49P earphones referenced to a Bruel and Kjaer DB-0138 HA-2 and NBS-9A acoustic coupler, respectively. The mean unadjusted and adjusted ER-3A thresholds were in very good agreement with the mean thresholds averaged over eight other studies. Further, the mean adjusted ER-3A thresholds were in very good agreement with the interim ANSI RETSPLs while the mean unadjusted ER-3A thresholds were slightly higher than the ISO RETSPLs. Also, mean ER-3A thresholds averaged over a nine study data base were in closer agreement with the interim ANSI than the ISO RETSPLS. Overall, the interim ANSI insert earphone RETSPLs were recommended for clinical use.
本研究的目的是获得ER-3A插入式耳机的正常听力阈值,为未来的标准化工作做出贡献,并与先前的研究以及美国国家标准学会(ANSI)和国际标准化组织(ISO)的临时插入式耳机参考等效阈值声压级(RETSPL)进行比较。使用分别参考Bruel和Kjaer DB-0138 HA-2和美国国家标准局(NBS)-9A声耦合器的ER-3A和TDH-49P耳机,对48名听力正常的受试者的每只耳朵在125至8000Hz频率范围内进行听力阈值测试。未调整和调整后的ER-3A阈值均值与其他八项研究的平均阈值非常吻合。此外,调整后的ER-3A阈值均值与ANSI临时RETSPL非常吻合,而未调整的ER-3A阈值均值略高于ISO RETSPL。同样,在九个研究数据库中平均得到的ER-3A阈值均值与ANSI临时值的吻合度比与ISO RETSPL的吻合度更高。总体而言,推荐将ANSI临时插入式耳机RETSPL用于临床。