Suppr超能文献

一种用于生物分子结构分析的紧凑型飞行时间质谱仪,采用激光解吸技术。

A compact time-of-flight mass spectrometer for the structural analysis of biological molecules using laser desorption.

作者信息

Cornish T, Cotter R J

机构信息

Department of Pharmacology and Molecular Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205.

出版信息

Rapid Commun Mass Spectrom. 1992 Apr;6(4):242-8. doi: 10.1002/rcm.1290060404.

Abstract

A compact laser-desorption time-of-flight mass spectrometer using a 600 ps nitrogen laser and 1 Gsample/s transient recorder is described. The instrument incorporates two electrically-isolatable, reflecting flight tubes designed for subsequent configuration of the mass spectrometer as a tandem instrument. In this first report, we compare mass resolution in the laser-desorption mass spectra of an organic dye, sinapinic and caffeic acid matrices, and several small peptides. For directly desorbed ions, peak widths are generally of the order of 11 to 13 ns, so that mass resolution increases with increasing mass. For peptide ions in the range of 500 to 1000 u, formed by matrix-assisted desorption, peak widths range from 9.5 to 17.6 ns and increase with mass. However, better than unit mass resolution (1600 to 2400) is maintained throughout this mass range. Mass measurement accuracy is better than 0.1 u using a single calibration peak and a prompt start trigger pulse. Prospects for extending mass range and resolution are discussed.

摘要

描述了一种使用600皮秒氮激光器和1吉采样每秒瞬态记录仪的紧凑型激光解吸飞行时间质谱仪。该仪器包含两个电隔离的反射飞行管,设计用于随后将质谱仪配置为串联仪器。在这份首次报告中,我们比较了有机染料、芥子酸和咖啡酸基质以及几种小肽的激光解吸质谱中的质量分辨率。对于直接解吸的离子,峰宽通常在11到13纳秒量级,因此质量分辨率随质量增加而提高。对于通过基质辅助解吸形成的500至1000 u范围内的肽离子,峰宽在9.5至17.6纳秒之间,并随质量增加。然而,在整个质量范围内保持了优于单位质量分辨率(1600至2400)。使用单个校准峰和快速启动触发脉冲时,质量测量精度优于0.1 u。讨论了扩展质量范围和分辨率的前景。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验