Yang Chih-Chio, Kumar Annamalai Senthil, Zen Jyh-Myng
Department of Chemistry, National Chung Hsing University, Taichung 402, Taiwan.
Anal Biochem. 2005 Mar 15;338(2):278-83. doi: 10.1016/j.ab.2004.12.015.
We report here a highly accurate and efficient method for blood lead analysis (BLA) through the use of a disposable electrode. A new type of mercury-plated preanodized screen-printed carbon electrode, together with a thallium(III) internal standard, simplifies calibration and gives easily quantifiable signals for accurate BLA. A preanodization procedure improves the preconcentration ability of the working electrode through the effect of lead (Pb) complexing with an electrogenerated surface functional group, mainly >C=O. Under optimized conditions, the ratio of the anodic stripping peak currents of Pb and Tl (i.e., i(Pb)/i(Tl)) is linear against [Pb] in the window of 1-300 ppb with a correlation coefficient and detection limit (signal-to-noise ratio=3) of 0.999 and 0.23 ppb, respectively. This approach was used to analyze blood samples (n=55) from workers at a local battery factory. The results are consistent with those obtained from graphite furnace atomic absorption spectroscopy and confirm the applicability of the proposed method.
我们在此报告一种通过使用一次性电极进行血铅分析(BLA)的高度准确且高效的方法。一种新型的镀汞预阳极氧化丝网印刷碳电极,结合铊(III)内标,简化了校准过程,并为准确的血铅分析提供易于量化的信号。预阳极氧化程序通过铅(Pb)与电生成的表面官能团(主要是>C=O)络合的作用,提高了工作电极的预富集能力。在优化条件下,Pb和Tl的阳极溶出峰电流之比(即i(Pb)/i(Tl))在1 - 300 ppb的范围内与[Pb]呈线性关系,相关系数和检测限(信噪比 = 3)分别为0.999和0.23 ppb。该方法用于分析当地一家电池厂工人的血样(n = 55)。结果与石墨炉原子吸收光谱法获得的结果一致,证实了所提出方法的适用性。