Schöfer H, Kotsianos D, Wirth S, Britsch S, Reiser M
Bayerisches Landesamt für Arbeitsschutz, Arbeitsmedizin und Sicherheitstechnik, München.
Radiologe. 2005 Mar;45(3):278-85. doi: 10.1007/s00117-005-1180-7.
Digital imaging in mammography is becoming more and more accepted using both computed (CR) and direct radiography (DR). These techniques will soon be used in screening programs. Therefore, quality assurance for this technique is indispensable. The relevance of the current regulations, such as EPOC and the German QS-RL was investigated. For the investigation, a breast phantom and appropriate software were developed. Both were tested using digital mammography systems from six manufactures. Quality assurance parameters (such as contrast to noise ratio and contrast resolution) were calculated from these data sets. The results should be considered in future standards for mammography (IEC respectively DIN). In addition, this type of test procedure is time saving and enables a reduction in test devices, i. e. in costs.