Digital imaging in mammography is becoming more and more accepted using both computed (CR) and direct radiography (DR). These techniques will soon be used in screening programs. Therefore, quality assurance for this technique is indispensable. The relevance of the current regulations, such as EPOC and the German QS-RL was investigated. For the investigation, a breast phantom and appropriate software were developed. Both were tested using digital mammography systems from six manufactures. Quality assurance parameters (such as contrast to noise ratio and contrast resolution) were calculated from these data sets. The results should be considered in future standards for mammography (IEC respectively DIN). In addition, this type of test procedure is time saving and enables a reduction in test devices, i. e. in costs.

摘要

乳腺X线摄影中的数字成像越来越多地被计算机X线摄影(CR)和直接数字化X线摄影(DR)所采用。这些技术很快将用于筛查项目。因此,该技术的质量保证不可或缺。对诸如EPOC和德国质量标准条例(QS-RL)等现行法规的相关性进行了调查。为进行该调查,开发了一个乳房模型和相应软件。两者均使用来自六个制造商的数字乳腺摄影系统进行测试。从这些数据集计算出质量保证参数(如对比度噪声比和对比分辨率)。这些结果应在未来的乳腺摄影标准(分别为国际电工委员会(IEC)标准和德国工业标准(DIN))中予以考虑。此外,这种测试程序节省时间,能够减少测试设备,即降低成本。