Okpara-Hofmann J, Knoll M, Dürr M, Schmitt B, Borneff-Lipp M
Institute of Hygiene, Medical Faculty, Martin-Luther-University Halle-Wittenberg, Johann-Andreas-Segner-Str. 12, D-06097 Halle/Saale, Germany.
J Hosp Infect. 2005 Apr;59(4):280-5. doi: 10.1016/j.jhin.2004.10.002.
This study compared the effectiveness of sterilizing four types of endoscope using different models of the Sterrad system (Sterrad 50, 100, 100S and 200). Sterilization levels meeting international requirements were attained in all cases with carriers inoculated with Geobacillus stearothermophilus spores. The endoscopes were tested in half cycles ('overkill'). This is the first study to compare the Sterrad models marketed to date in terms of effective sterilization of endoscopes with narrow lumens.
本研究比较了使用不同型号的Sterrad系统(Sterrad 50、100、100S和200)对四种类型的内窥镜进行灭菌的效果。在所有接种嗜热栖热放线菌孢子的载体的情况下,均达到了符合国际要求的灭菌水平。内窥镜在半周期(“过度杀灭”)中进行了测试。这是第一项比较迄今为止上市的Sterrad各型号在有效灭菌窄腔道内窥镜方面的研究。