Tang Maolong, Shekhar Raj, Huang David
IEEE Trans Med Imaging. 2005 Mar;24(3):424-8. doi: 10.1109/tmi.2004.843192.
Corneal topography is used to measure the anterior surface of the cornea. It is conventionally represented as radial slope, radial curvature, and elevation. In this paper, we introduce the application of mean curvature mapping as an alternative representation of the corneal topography. The purpose is to improve the detection of keratoconus and other diseases characterized by local increase in corneal curvature. Both simulated keratoconic cornea and real keratoconus data exported from the corneal topography system were analyzed. Four representations of corneal topography were generated and compared. It was found that mean curvature mapping provided the most precise cone location in simulated keratoconus. In both actual and simulated keratoconus cases, the appearance of the cone-like distortion is more consistent on mean curvature maps. Mean curvature mapping may improve the detection and localization of corneal shape abnormalities.
角膜地形图用于测量角膜前表面。传统上它以径向斜率、径向曲率和高度来表示。在本文中,我们介绍了平均曲率映射作为角膜地形图的一种替代表示方法的应用。目的是提高圆锥角膜和其他以角膜曲率局部增加为特征的疾病的检测。分析了模拟圆锥角膜和从角膜地形图系统导出的真实圆锥角膜数据。生成并比较了角膜地形图的四种表示方法。发现在模拟圆锥角膜中,平均曲率映射提供了最精确的圆锥位置。在实际和模拟圆锥角膜病例中,平均曲率图上的圆锥样畸变外观更一致。平均曲率映射可能会改善角膜形状异常的检测和定位。