Liu Yingjie, Allaire Paul, Wood Houston, Olsen Don
Department of Mechanical & Aerospace Engineering, University of Virginia, Charlottesville, VA 22903, USA.
Artif Organs. 2005 Apr;29(4):341-5. doi: 10.1111/j.1525-1594.2005.29058.x.
A mock circulatory loop, which simulates the human circulatory system, is needed to bench test the various versions of continuous flow (CF) left ventricular assist devices (LVADs). This article describes the design and initial testing of such a loop. The loop consists of: (1) pulsatile left and right cardiac simulators; (2) air/water tanks to model the venous and arterial compliances; (3) tygon tubes to model the venous, arterial, and other system flow resistances; and (4) a tuning clamp to model the variation in system resistance characteristics under different cardiac pressure/flow conditions. Several loop measurements were carried out without an LVAD to verify the cardiovascular modeling of a healthy person in sleep, rest, and physical activity, and in different pathological states, and compared to the data found in the literature to validate the loop performance prior to LVAD testing.
需要一个模拟人体循环系统的模拟循环回路来对各种版本的连续流(CF)左心室辅助装置(LVAD)进行台架测试。本文描述了这种回路的设计和初始测试。该回路包括:(1)搏动性左、右心脏模拟器;(2)用于模拟静脉和动脉顺应性的空气/水箱;(3)用于模拟静脉、动脉和其他系统流动阻力的泰根管;以及(4)用于模拟不同心压/流量条件下系统阻力特性变化的调谐夹。在没有LVAD的情况下进行了几次回路测量,以验证健康人在睡眠、休息和身体活动以及不同病理状态下的心血管模型,并与文献中的数据进行比较,以在LVAD测试之前验证回路性能。