The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert.

作者信息

McCormack Stephen, Diffey Jennifer, Morgan Andrew

机构信息

Department of Medical Physics, Cookridge Hospital, Hospital Lane, Leeds LS16 6QB, England.

出版信息

Med Phys. 2005 Feb;32(2):483-7. doi: 10.1118/1.1852792.

Abstract

The use of rigid carbon fiber couch inserts in radiotherapy treatment couches is a well-established method of reducing patient set-up errors associated with couch sag. Several published studies have described such inserts as radiotranslucent with negligible attenuation of the radiation field. Most of these studies were conducted with the radiation field normally incident on the couch and there appears to be no evidence in the literature of the effect of the gantry angle on the extent of beam attenuation by the carbon fiber insert alone during external beam radiotherapy. In this study we examined the magnitude of this effect over a range of posterior oblique gantry angles using a cylindrical solid water phantom containing an ionization chamber placed isocentrically. It was found that a 6 MV photon beam, field size 10 x 5 cm, was attenuated significantly as the gantry angle approached the plane of the couch, from 2% at normal incidence and reaching 9% attenuation at angle of incidence 70 degrees. This could have serious implications regarding dose to the treatment volume for treatments requiring posterior oblique angles of incidence with a possible correction factor necessary in monitor unit calculations.

摘要

在放射治疗治疗床上使用刚性碳纤维床板插入物是一种成熟的减少与床体下垂相关的患者摆位误差的方法。多项已发表的研究将此类插入物描述为射线半透明,对辐射野的衰减可忽略不计。这些研究大多是在辐射野垂直入射到治疗床的情况下进行的,而且文献中似乎没有关于在体外放射治疗期间,仅碳纤维插入物对机架角度对射束衰减程度影响的证据。在本研究中,我们使用一个包含等中心放置的电离室的圆柱形固体水模体,在一系列后斜机架角度范围内研究了这种影响的大小。结果发现,对于6兆伏光子束,射野尺寸为10×5厘米,当机架角度接近治疗床平面时,射束衰减显著,从垂直入射时的2%增加到入射角为70度时的9%衰减。这对于需要后斜入射角的治疗中治疗体积的剂量可能有严重影响,在监测单位计算中可能需要一个校正因子。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验