Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants.

作者信息

Rieker Claude B, Schön Rolf, Konrad Reto, Liebentritt Gernot, Gnepf Patric, Shen Ming, Roberts Paul, Grigoris Peter

机构信息

Tribology-113 955, Zimmer Europe GmbH, Sulzer Allee 8, CH 8404 Winterthur, Switzerland.

出版信息

Orthop Clin North Am. 2005 Apr;36(2):135-42, vii. doi: 10.1016/j.ocl.2005.02.004.

Abstract

Large-diameter metal-on-metal articulations may provide an opportunity for wear reduction in total hip implants because earlier studies have shown that the formation of a fluid film that completely separates the bearing surfaces is theoretically possible. In such a lubrication mode and under ideal conditions, there is theoretically no amount of wear. Studies have suggested that the two primary parameters controlling the lubrication mode are the diameter and the clearance of the articulation. The goal of the present study was to experimentally investigate the influence of these two parameters on the wear behavior of large-diameter metal-on-metal articulations pertaining to resurfacing hip implants. The results of this in vitro investigation showed that longer running-in periods and higher amounts of running-in wear were associated with larger clearances.