Evaluation of film and thermoluminescent dosimetry of high-energy electron beams in heterogeneous phantoms.

作者信息

el-Khatib E, Antolak J, Scrimger J

机构信息

Division of Medical Physics, British Columbia Cancer Agency, Vancouver, Canada.

出版信息

Med Phys. 1992 Mar-Apr;19(2):317-23. doi: 10.1118/1.596861.

Abstract

Film and thermoluminescent dosimetry (TLD) are investigated in heterogeneous phantoms irradiated by high-energy electron beams. Both film and TLD are practical dosimeters for multiple and moving beam radiotherapy. The accuracy and precision of these dosimeters for radiation dose measurements in homogeneous water-equivalent phantoms has been discussed in the literature. However, film and TLD are often used for dose measurements in heterogeneous phantoms. In those situations perturbations are produced which are related to the density and atomic number of the phantom material and the physical size and orientation of the dosimeter. In our experiments the relative dose measurements in homogeneous phantoms were the same regardless of dosimeter or dosimeter orientation. However, significant differences were observed between the dose measurements within the inhomogeneity. These differences were influenced by the type and orientation of the dosimeter in addition to the properties of the heterogeneity. These differences could be reproduced with Monte Carlo calculations and modeling of the experimental conditions.