Farer Leslie J
Schering-Plough Research Institute, Analytical Development, 2000 Galloping Hill Rd, Kenilworth, NJ 07033, USA.
J AOAC Int. 2005 Mar-Apr;88(2):462-7.
A new method was developed for the quantitation of emamectin benzoate in medicated fish feed at levels between 1 and 30 ppm. The new procedure, based on a previously reported assay, consists of a wet methanolic extraction of ground feed, followed by solid-phase extraction and injection onto a gradient liquid chromatographic system. A multisite study involving 3 laboratories (the developing laboratory and 2 independent laboratories) was performed to evaluate precision, recovery, linearity, and sensitivity. Mean recove;ries for triplicate analyses at 3 levels, performed by 2 analysts per laboratory, were between 89 and 97%, with coefficients of variation ranging from 1.6 to 8.6%. Coefficients of determination (r2) obtained from the plotted data were > or =0.993. The precision of the method, determined from 6 replicate preparations from the same batch of medicated feed assayed in 3 separate trials per laboratory, was between 0.6 and 5.8%. The quantitation limit was established at 0.5 ppm. Specificity and robustness studies were performed by the developing laboratory.