犬脉搏血氧饱和度测定法准确性的评估。

Evaluation of accuracy of pulse oximetry in dogs.

作者信息

Jacobson J D, Miller M W, Matthews N S, Hartsfield S M, Knauer K W

机构信息

Department of Small Animal Medicine and Surgery, College of Veterinary Medicine, Texas A&M University, College Station 77843.

出版信息

Am J Vet Res. 1992 Apr;53(4):537-40.

PMID:1586024
Abstract

The accuracy of a pulse oximeter was evaluated over a wide range of arterial oxygen and carbon dioxide tensions, using 2 probes (finger probe and ear probe) and 2 monitoring sites (tongue and tail) in anesthetized dogs. The arterial oxygen saturation of hemoglobin (SaO2) measured directly with a multiwavelength spectrophotometer was compared with saturation estimated by pulse oximetry (SpO2). Linear regression analysis of the pooled data from 399 simultaneous measurements of SpO2 and SaO2 indicated a highly significant correlation of SpO2 with SaO2 (r = 0.97; P less than or equal to 0.0001). Although the mean difference (+/- SD) between SpO2 and SaO2 for pooled data was small (-0.06 +/- 6.8%), SpO2 tended to underestimate high SaO2 values (greater than or equal to 70%) and to overestimate low SaO2 values (less than 70%). When SaO2 values were greater than or equal to 70%, the ear probe applied to the tail was less accurate (produced a significantly greater SpO2-SaO2 difference) than the ear probe on the tongue, or the finger probe at either site. When SaO2 values were less than or equal to 50%, the finger probe applied at the tail was more accurate (produced significantly smaller SpO2-SaO2 differences) than the ear probe at either site. When SaO2 values were less than or equal to 70%, high arterial carbon dioxide tension (greater than or equal to 60 mm of Hg) was associated with greater overestimation of SaO2.

摘要

使用2个探头(手指探头和耳部探头)和2个监测部位(舌头和尾巴),在麻醉犬身上,于广泛的动脉氧分压和二氧化碳分压范围内评估脉搏血氧仪的准确性。将用多波长分光光度计直接测量的血红蛋白动脉血氧饱和度(SaO2)与通过脉搏血氧测定法估算的饱和度(SpO2)进行比较。对399次同时测量SpO2和SaO2的汇总数据进行线性回归分析表明,SpO2与SaO2具有高度显著的相关性(r = 0.97;P≤0.0001)。尽管汇总数据中SpO2和SaO2之间的平均差异(±标准差)很小(-0.06±6.8%),但SpO2往往会低估高SaO2值(≥70%),高估低SaO2值(<70%)。当SaO2值≥70%时,应用于尾巴的耳部探头比应用于舌头的耳部探头或任一部位的手指探头准确性更低(SpO2 - SaO2差异显著更大)。当SaO2值≤50%时,应用于尾巴的手指探头比任一部位的耳部探头准确性更高(SpO2 - SaO2差异显著更小)。当SaO2值≤70%时,高动脉二氧化碳分压(≥60 mmHg)与对SaO2的更大高估有关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验