Problematic or dysfunctional hemodialysis (HD) catheters are routinely reversed to achieve adequate blood flow for dialysis delivery. The purpose of the study is to determine the effect of varying blood pump speed (Qb) on access recirculation (AR), and urea clearance (K) in dysfunctional catheters in the normal and reversed positions.
Nineteen HD patients with tunneled cuffed catheters (5 functional and 14 dysfunctional catheters) were included; dysfunctional catheters are defined as the inability to attain a Qb of 300 mL/min or greater on 2 consecutive HD runs. AR and K measurements were obtained systematically for each catheter in the normal and reversed positions at increasing Qbs. K was measured using the ionic dialysance technique.
In functional catheters, AR in the normal position was 0% and increased to 15% +/- 13% when reversed. Dysfunctional catheters had a greater AR of 25% +/- 16% when reversed. In functional catheters, there was no evidence of an increase in AR with increasing Qb irrespective of position. Similarly, there was no relationship between increasing AR and greater Qbs (r 2 = 0.10) in dysfunctional catheters. In dysfunctional catheters, when reversed, mean K increased from 128 +/- 10 mL/min at a Qb of 200 mL/min to 157 +/- 38 mL/min at maximal Qb (P < 0.05).
We show that at increasing Qbs, K is improved in both functional and dysfunctional catheters. Data from the study are used to describe a nomogram to determine minimum Qb for a dysfunctional catheter in reversed position to maximize K.
有问题或功能失调的血液透析(HD)导管通常会被翻转,以实现足够的血流量用于透析。本研究的目的是确定不同的血泵速度(Qb)对功能失调导管在正常和翻转位置时的通路再循环(AR)和尿素清除率(K)的影响。
纳入19例带隧道带 cuff 的HD患者(5根功能正常的导管和14根功能失调的导管);功能失调的导管定义为在连续2次HD运行中无法达到300 mL/分钟或更高的Qb。在正常和翻转位置,随着Qb增加,系统地测量每根导管的AR和K。使用离子透析技术测量K。
在功能正常的导管中,正常位置的AR为0%,翻转时增加到15%±13%。功能失调的导管翻转时AR更高,为25%±16%。在功能正常的导管中,无论位置如何,均无证据表明随着Qb增加AR会升高。同样,在功能失调的导管中,AR增加与更高的Qb之间也没有关系(r² = 0.10)。在功能失调的导管中,翻转时,平均K从Qb为200 mL/分钟时的128±10 mL/分钟增加到最大Qb时的157±38 mL/分钟(P < 0.05)。
我们表明,随着Qb增加,功能正常和功能失调的导管的K均得到改善。该研究的数据用于描述一个列线图,以确定功能失调导管在翻转位置时的最小Qb,以使K最大化。