An improved outdoor calibration procedure for broadband ultraviolet radiometers.

作者信息

Cancillo M L, Serrano A, Antón M, García J A, Vilaplana J M, de la Morena B

机构信息

Departamento de Física, Universidad de Extremadura, 06071 Badajoz, Spain.

出版信息

Photochem Photobiol. 2005 Jul-Aug;81(4):860-5. doi: 10.1562/2005-01-12-RA-412.

Abstract

This article aims at improving the broadband ultraviolet radiometer's calibration methodology. For this goal, three broadband radiometers are calibrated against a spectrophotometer of reference. Three different one-step calibration models are tested: ratio, first order and second order. The latter is proposed in order to adequately reproduce the high dependence on the solar zenith angle shown by the other two models and, therefore, to improve the calibration performance at high solar elevations. The proposed new second-order model requires no additional information and, thus, keeps the operational character of the one-step methodology. The models are compared in terms of their root mean square error and the most qualified is subsequently validated by comparing its predictions with the spectrophotometer measurements within an independent validation data subset. Results show that the best calibration is achieved by the second-order model, with a mean bias error and mean absolute bias error lower than 2.2 and 6.7%, respectively.