Reinke M
Department of Medicine IV, University of Erlangen-Nürnberg, Federal Republic of Germany.
Am J Physiol. 1992 May;262(5 Pt 1):E658-62. doi: 10.1152/ajpendo.1992.262.5.E658.
11-Dehydrothromboxane B2 (11-DHTxB2) concentrations are believed to reflect levels of the in vivo synthesis and release of thromboxane A2. In a specific enzyme-linked immunosorbent assay (ELISA) a monoclonal antibody (MAB) against 11-DHTxB2 (MAB-1E-DHTBR1) recognizes the acyclic form of 11-DHTxB2, as found in the basic range of pH, with a detection limit of 4.6 pg/sample and a binding affinity of 6.1 x 10(9) l/mol. Negligible cross-reactivity was found for thromboxane B2 (0.05%), 2,3-dinor-thromboxane B2 (0.06%), and prostaglandin D2 (0.08%). Validity of the assay was confirmed by a good correlation between radioimmunoassay and ELISA (r = 0.972). Recovery experiments showed an accuracy of r = 0.982. Measurements of 11-DHTxB2 in human serum and urine samples demonstrated the practical applicability of the MAB in ELISA. With the use of different clotting times, the serum level of 11-DHTxB2 ranged from 0.8-1.3 ng/ml (30 min) to 24.1-47.9 ng/ml (4 h). After administration of aspirin the 11-DHTxB2 level of human urine declined from 3.9-5.4 ng/ml to 0.4-1.6 ng/ml after 6 h.
11-脱氢血栓素B2(11-DHTxB2)的浓度被认为反映了血栓素A2在体内的合成和释放水平。在一种特定的酶联免疫吸附测定(ELISA)中,一种针对11-DHTxB2的单克隆抗体(MAB-1E-DHTBR1)可识别在pH值基本范围内发现的11-DHTxB2的无环形式,检测限为4.6 pg/样本,结合亲和力为6.1×10⁹ l/mol。发现血栓素B2(0.05%)、2,3-二去甲血栓素B2(0.06%)和前列腺素D2(0.08%)的交叉反应可忽略不计。放射免疫测定和ELISA之间的良好相关性(r = 0.972)证实了该测定的有效性。回收率实验显示准确度为r = 0.982。对人血清和尿液样本中11-DHTxB2的测量证明了该单克隆抗体在ELISA中的实际适用性。使用不同的凝血时间,11-DHTxB2的血清水平范围为0.8 - 1.3 ng/ml(30分钟)至24.1 - 47.9 ng/ml(4小时)。服用阿司匹林后,人尿液中11-DHTxB2的水平在6小时后从3.9 - 5.4 ng/ml降至0.4 - 1.6 ng/ml。