A comparison of the aspirating abilities of re-usable and partly disposable dental cartridge syringes in vitro.

作者信息

Meechan J G, Ramacciato J C, McCabe J F

机构信息

Department of Oral Surgery, School of Dental Sciences, University of Newcastle upon Tyne, Newcastle upon Tyne, England NE2 4BW, UK.

出版信息

J Dent. 2006 Jan;34(1):41-7. doi: 10.1016/j.jdent.2005.03.003.

Abstract

OBJECTIVES

To compare the ability of re-usable and partly disposable dental cartridge syringes to aspirate in vitro.

METHODS

A laboratory investigation comparing two re-usable (Astra and Rotor) and one partly disposable (Ultra Safety-plus) dental cartridge syringes. Forces needed to move the cartridge plunger and forces required to produce aspiration of Bonney's Blue dye were measured using an Instron testing machine.

RESULTS

There were significant differences in the forces needed to produce aspiration in the different systems (F = 194, p<0.001). The Astra system needed the least force to produce aspiration. The forces required to aspirate in the Rotor and Ultra Safety-plus systems did not differ. Each system aspirated effectively in vitro at the initial aspiration manoeuvre. The forces available for subsequent aspiration attempts varied with the speed of cartridge plunger movement. The volume of local anaesthetic cartridge used did not affect the force needed to produce aspiration in the Ultra Safety-plus syringe. The method of testing free-flow of solution through the needle affected the force generated at subsequent plunger movements.

CONCLUSIONS

The re-usable and partly disposable syringes investigated aspirated effectively at the initial aspiration manoeuvre. The force available to produce aspiration at subsequent attempts varies with the speed of injection. The method of testing free-flow of solution through the needle in a loaded syringe affects the force available at the first aspiration manoeuvre.

摘要

目的

比较可重复使用和部分一次性牙科卡筒注射器的体外抽吸能力。

方法

进行一项实验室研究,比较两种可重复使用的(Astra和Rotor)和一种部分一次性的(Ultra Safety-plus)牙科卡筒注射器。使用英斯特朗测试机测量推动卡筒活塞所需的力以及产生Bonney氏蓝染料抽吸所需的力。

结果

不同系统产生抽吸所需的力存在显著差异(F = 194,p<0.001)。Astra系统产生抽吸所需的力最小。Rotor系统和Ultra Safety-plus系统抽吸所需的力没有差异。每个系统在初始抽吸操作时在体外均能有效抽吸。后续抽吸尝试可用的力随卡筒活塞移动速度而变化。Ultra Safety-plus注射器中使用的局部麻醉剂卡筒的体积不影响产生抽吸所需的力。测试溶液通过针头自由流动的方法影响后续活塞移动时产生的力。

结论

所研究的可重复使用和部分一次性注射器在初始抽吸操作时能有效抽吸。后续尝试产生抽吸可用的力随注射速度而变化。测试已装载注射器中溶液通过针头自由流动的方法影响首次抽吸操作时可用的力。