Acad Radiol. 2005 Jun;12(6):771-5. doi: 10.1016/j.acra.2005.02.013.
To determine the optimal gadolinium concentration for catheter-directed coronary magnetic resonance angiography (MRA) using magnetization-prepared steady-state free-precession (SSFP) in swine.
In six pigs, we performed real-time MR imaging-guided coronary artery catheterization using a 1.5 T MR scanner. For catheter-directed coronary MRA, we injected 3-4 mL of dilute Gd at 1 mL/second for each tested concentration (4%, 8%, 10%, and 12% Gd). Eleven images per concentration were acquired using electrocardiographic-triggered, magnetization-prepared two-dimensional (2D) projection SSFP. We compared mean relative signal-to-noise ratio (SNR) values for each concentration using two-way analysis of variance.
The targeted coronary artery was catheterized under real-time MR guidance in all pigs. Magnetization-prepared 2D projection SSFP successfully depicted the coronary arteries in all 44 injections. Mean relative SNR (+/- standard error) was 7.2 +/- 0.49 for 4%, 8.8 +/- 0.47 for 8%, 9.5 +/- 0.38 for 10%, and 8.8 +/- 0.41 for 12%. Injections of 4% dilute gadolinium yielded significantly less relative SNR than the other tested concentrations (P < .05). There were no statistically significant differences between the remaining concentrations.
For catheter-directed contrast-enhanced coronary MRA, the ideal gadolinium concentration should maximize relative SNR and limit the total gadolinium dose. Using these criteria, of those concentrations we tested in the swine model, 8% injected gadolinium was superior for catheter-directed SSFP coronary MRA.
确定猪体内采用磁化准备稳态自由进动(SSFP)技术进行导管引导冠状动脉磁共振血管造影(MRA)的最佳钆浓度。
对6头猪使用1.5T磁共振扫描仪进行实时磁共振成像引导下的冠状动脉插管。对于导管引导冠状动脉MRA,每种测试浓度(4%、8%、10%和12%钆)以1mL/秒的速度注射3 - 4mL稀释钆。使用心电图触发的、磁化准备二维(2D)投影SSFP采集每种浓度下的11幅图像。我们采用双向方差分析比较每种浓度下的平均相对信噪比(SNR)值。
所有猪均在实时磁共振引导下完成目标冠状动脉插管。磁化准备2D投影SSFP在所有44次注射中均成功显示冠状动脉。4%浓度时平均相对SNR(±标准误差)为7.2±0.49,8%时为8.8±0.47,10%时为9.5±0.38,12%时为8.8±0.41。注射4%稀释钆产生的相对SNR显著低于其他测试浓度(P <.05)。其余浓度之间无统计学显著差异。
对于导管引导对比增强冠状动脉MRA,理想的钆浓度应使相对SNR最大化并限制总钆剂量。根据这些标准,在我们猪模型中测试的浓度里,8%注射钆对于导管引导SSFP冠状动脉MRA是最优的。