Use of progressive glaucomatous optic disk change as the reference standard for evaluation of diagnostic tests in glaucoma.

作者信息

Medeiros Felipe A, Zangwill Linda M, Bowd Christopher, Sample Pamela A, Weinreb Robert N

机构信息

Hamilton Glaucoma Center and Department of Ophthalmology, University of California-San Diego, La Jolla, CA 92093-0946, USA.

出版信息

Am J Ophthalmol. 2005 Jun;139(6):1010-8. doi: 10.1016/j.ajo.2005.01.003.

Abstract

PURPOSE

Progressive change in optic disk appearance has been suggested as the best available reference standard for glaucoma diagnosis. We evaluated the accuracy of scanning laser polarimetry (SLP) with variable corneal compensation to discriminate patients with documented progressive optic disk change from healthy subjects.

DESIGN

Observational case-control study.

METHODS

Glaucoma patients with documented evidence of progressive optic disk change as assessed by stereoscopic photographs were selected. Healthy control subjects had normal visual fields and normal clinical examination. All subjects underwent imaging with the GDx VCC scanning laser polarimeter. Receiver-operating characteristic (ROC) curves, sensitivities at fixed specificities, and likelihood ratios were calculated.

RESULTS

Sixty-five eyes were healthy control patients. Seventy-one eyes had progressive optic disk change, of which 41 (58%) had abnormal visual field tests closest to the imaging date (perimetric glaucomatous optic neuropathy [GON]), and 30 (42%) eyes had normal visual fields (preperimetric GON). The GDx VCC parameter nerve fiber indicator performed best for discrimination between perimetric GON and normal eyes (AUC = 0.94; sensitivity = 93% for specificity > or =80%) as well as for the discrimination between preperimetric GON and normal eyes (AUC = 0.89; sensitivity = 83% for specificity > or =80%).

CONCLUSION

SLP differentiated patients with documented progressive optic disk change from normal subjects and detected abnormalities in patients with confirmed diagnosis of preperimetric GON, suggesting that it may be useful for early glaucoma diagnosis. We suggest that progressive optic disk change should be used in the evaluation of diagnostic accuracy of imaging instruments for glaucoma.

摘要

目的

视盘外观的渐进性变化被认为是青光眼诊断的最佳可用参考标准。我们评估了具有可变角膜补偿的扫描激光偏振仪(SLP)区分有记录的视盘渐进性变化患者与健康受试者的准确性。

设计

观察性病例对照研究。

方法

选择通过立体照片评估有视盘渐进性变化记录证据的青光眼患者。健康对照受试者视野正常且临床检查正常。所有受试者均使用GDx VCC扫描激光偏振仪进行成像。计算受试者操作特征(ROC)曲线、固定特异性下的敏感性和似然比。

结果

65只眼为健康对照患者。71只眼有视盘渐进性变化,其中41只眼(58%)在最接近成像日期的视野检查中异常(周边视野青光眼性视神经病变[GON]),30只眼(42%)视野正常(周边视野前GON)。GDx VCC参数神经纤维指标在区分周边视野GON与正常眼方面表现最佳(曲线下面积[AUC]=0.94;特异性≥80%时敏感性=93%),在区分周边视野前GON与正常眼方面也表现最佳(AUC = 0.89;特异性≥80%时敏感性=83%)。

结论

SLP能够区分有记录的视盘渐进性变化患者与正常受试者,并检测出确诊为周边视野前GON患者的异常情况,这表明它可能有助于青光眼的早期诊断。我们建议在评估青光眼成像仪器的诊断准确性时应采用视盘渐进性变化。