Jönsson B, Laurent C, Eneling M, Skau T, Lindberg L-G
Department of Cardiovascular Surgery and Anaesthesia, University Hospital, Linköping, Sweden.
Eur J Vasc Endovasc Surg. 2005 Oct;30(4):395-401. doi: 10.1016/j.ejvs.2005.05.012.
To evaluate a new technique using a photoplethysmographic (PPG) probe for automatic ankle pressure measurements.
Comparative study on two techniques for ankle pressure measurement.
University hospital.
Thirty-five patients with leg arterial disease and eight healthy volunteers. Ankle-brachial indices (ABPI) were measured using conventional CW Doppler technique and PPG-based prototype equipment for the ankle pressure recordings.
ABPIs calculated from CW Doppler and PPG ankle pressure measurements. The PPG signals were analysed both by visual judgement and by a software based, automatic algorithm.
The mean difference between ABPIs calculated from CW Doppler recordings and PPG (visual analysis) was -0.01 (limits of agreement (+/-two standard deviations) +0.16 to -0.19). The correlation coefficient was 0.93. When the algorithm was used, the mean difference (CW Doppler-PPG) was 0.05 (limits of agreement 0.28 to -0.18, r=0.89).
The PPG method is a promising technique with an inherent potential for automatisation of the ankle pressure measurements, thereby reducing the observer-dependency in ABPI recordings.