使用不同阳极/滤过器组合的两台数字乳腺摄影设备平均腺体剂量的个体内比较。

Intra-individual comparison of average glandular dose of two digital mammography units using different anode/filter combinations.

作者信息

Engelken Florian J F, Meyer Henning, Juran Ralf, Bick Ulrich, Fallenberg Eva, Diekmann Felix

机构信息

Institut für Radiologie, Charité Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany.

出版信息

Acad Radiol. 2009 Oct;16(10):1272-80. doi: 10.1016/j.acra.2009.05.012. Epub 2009 Jul 25.

Abstract

RATIONALE AND OBJECTIVES

The aim of this study was to compare average glandular dose (AGD) in two full-field digital mammography units using different anode/filter combinations.

MATERIALS AND METHODS

Mammographies of 50 consecutive patients on a mammography system using a tungsten/rhodium (W/Rh) anode/filter combination were retrospectively compared to prior examinations on a different mammography unit using combinations of Molybdenum (Mo) and Rhodium (Rh). To exclude effects of increasing patient age, two prior examinations within 5 years were used. Both views of one breast were chosen for analysis. AGD was recorded as stated by each mammography unit. Accuracy of stated AGD and contrast-detail resolution were assessed using different breast phantoms.

RESULTS

The mean AGDs from the examinations using W/Rh were 0.95 mGy and 1.01 mGy for craniocaudal (CC) and mediolateral oblique (MLO) views compared to 1.51 mGy and 1.54 mGy, respectively, using combinations of Mo and Rh (P < .001). Relative reduction of AGD was independent of breast thickness but decreased with increasing breast density (partial correlation coefficient of 0.46, P < .005 and 0.57, P < .001, for CC and MLO views, respectively). Low-level contrast resolution was equal in both units using standard acquisition parameters.

CONCLUSION

In clinical mammographies, higher energy beam spectra obtained using W/Rh anode/filter combinations may significantly contribute to lowering AGD compared to Mo/Mo, Mo/Rh, and Rh/Rh in breasts that are not extremely dense.

摘要

原理与目的

本研究旨在比较使用不同阳极/滤过组合的两种全视野数字化乳腺摄影设备的平均腺体剂量(AGD)。

材料与方法

回顾性比较了在一台使用钨/铑(W/Rh)阳极/滤过组合的乳腺摄影系统上对50例连续患者进行的乳腺摄影检查结果与在另一台使用钼(Mo)和铑(Rh)不同组合的乳腺摄影设备上进行的先前检查结果。为排除患者年龄增长的影响,选取了5年内的两次先前检查。选择一侧乳房的两个投照体位进行分析。各乳腺摄影设备记录的AGD值。使用不同的乳房模型评估记录的AGD值的准确性和对比细节分辨力。

结果

使用W/Rh进行检查时,头尾位(CC)和内外斜位(MLO)投照的平均AGD分别为0.95 mGy和1.01 mGy,而使用Mo和Rh组合时分别为1.51 mGy和1.54 mGy(P <.001)。AGD的相对降低与乳房厚度无关,但随乳房密度增加而降低(CC和MLO投照的偏相关系数分别为0.46,P <.005和0.57,P <.001)。使用标准采集参数时,两个设备的低对比度分辨力相同。

结论

在临床乳腺摄影中,对于密度不是极高的乳房,使用W/Rh阳极/滤过组合获得的较高能量束光谱可能显著有助于降低AGD,相比Mo/Mo、Mo/Rh和Rh/Rh组合。

相似文献

5
Patient dose in digital mammography.数字化乳腺摄影中的患者剂量。
Med Phys. 2004 Sep;31(9):2471-9. doi: 10.1118/1.1784591.
9
A search for optimal x-ray spectra in iodine contrast media mammography.碘对比剂乳腺摄影中最佳X线光谱的探索。
Phys Med Biol. 2005 Jul 7;50(13):3143-52. doi: 10.1088/0031-9155/50/13/012. Epub 2005 Jun 22.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验