放疗期间使用超声进行实时跟踪的可行性。
Feasibility of using ultrasound for real-time tracking during radiotherapy.
作者信息
Hsu A, Miller N R, Evans P M, Bamber J C, Webb S
机构信息
Joint Department of Physics, Institute of Cancer Research and Royal Marsden NHS Trust, Sutton, Surrey, SM2 5PT United Kingdom.
出版信息
Med Phys. 2005 Jun;32(6):1500-12. doi: 10.1118/1.1915934.
This study was designed to examine the feasibility of utilizing transabdominal ultrasound for real-time monitoring of target motion during a radiotherapy fraction. A clinical Acuson 128/XP ultrasound scanner was used to image various stationary and moving phantoms while an Elekta SL25 linear accelerator radiotherapy treatment machine was operating. The ultrasound transducer was positioned to image from the outer edge of the treatment field at all times. Images were acquired to videotape and analyzed using in-house motion tracking algorithms to determine the effect of the SL25 on the quality of the displacement measurements. To determine the effect on the dosimetry of the presence of the transducer, dose distributions were examined using thermoluminescent dosimeters loaded into an Alderson Rando phantom and exposed to a 10 x 10 cm2 treatment field with and without the ultrasound transducer mounted 2.5 cm outside the field edge. The ultrasound images acquired a periodic noise that was shown to occur at the pulsing frequency of the treatment machine. Images of moving tissue were analyzed and the standard deviation on the displacement estimates within the tissue was identical with the SL25 on and off. This implies that the periodic noise did not significantly degrade the precision of the tracking algorithm (which was better than 0.01 mm). The presence of the transducer at the surface of the phantom presented only a 2.6% change to the dose distribution to the volume of the phantom. The feasibility of ultrasonic motion tracking during radiotherapy treatment is demonstrated. This presents the possibility of developing a noninvasive, real-time and low-cost method of tracking target motion during a treatment fraction.
本研究旨在探讨利用经腹超声在放射治疗分次过程中实时监测靶区运动的可行性。在Elekta SL25直线加速器放射治疗机运行时,使用临床用的Acuson 128/XP超声扫描仪对各种静止和运动的体模进行成像。超声换能器始终定位在从治疗野的外边缘进行成像。采集图像并录制到录像带上,使用内部运动跟踪算法进行分析,以确定SL25对位移测量质量的影响。为了确定换能器的存在对剂量测定的影响,使用装入Alderson Rando体模的热释光剂量计,在有和没有将超声换能器安装在野边缘外2.5 cm处的情况下,对10×10 cm2的治疗野进行照射,检查剂量分布。采集的超声图像出现了一种周期性噪声,该噪声显示在治疗机的脉冲频率处出现。对运动组织的图像进行分析,组织内位移估计值的标准差在SL25开启和关闭时相同。这意味着周期性噪声并未显著降低跟踪算法的精度(优于0.01 mm)。换能器在体模表面的存在对体模体积的剂量分布仅产生了2.6%的变化。证明了放射治疗期间超声运动跟踪的可行性。这为开发一种在治疗分次过程中跟踪靶区运动的非侵入性、实时且低成本的方法提供了可能性。