Debrun Dominique, Thérain Fabienne, Nguyen Long-Dang, Léger Christophe P, Visser Jacco J N, Busemann-Sokole Ellinor
Service de Médecine Nucléaire, Centre Hospitalier Régional d' Orléans (CHRO), 14 Avenue de 1'Hôpital, BP 6709, 45067, Orléans Cedex 2, France.
Int J Cardiovasc Imaging. 2005 Apr-Jun;21(2-3):239-47; discussion 249-51. doi: 10.1007/s10554-004-4014-1.
A dynamic cardiac phantom was used as a reference to compare volumes measured with gated SPECT and 4D echocardiography.
Gated SPECT data were acquired with a standard single-head gamma camera, and the volume reconstructions were carried out using the Mirage software by Segami. 4D echocardiography used a new prototype of rotating scan head to acquire ultrasound images during a cardiac cycle, used to reconstruct the volume deformations as a function of time. End-diastolic volume, end-systolic volume, and ejection fraction were measured using both gated SPECT and 4D echocardiography.
The results obtained showed a good correlation between volumes measured with the two modalities, but a slight overestimation of volumes with gated SPECT. The influence of filtering and pixel size parameters on the measured volumes was quantified for gated SPECT, in order to correct the overestimation.
The agreement between gated SPECT (after correction) and 4D echocardiography confirmed the relevance of the comparisons. This study was an initial step before conducting clinical trials to compare exhaustively left ventricular volumes obtained with the two modalities.
使用动态心脏模型作为参考,比较门控单光子发射计算机断层扫描(gated SPECT)和四维超声心动图测量的容积。
使用标准单探头伽马相机采集门控SPECT数据,并使用Segami公司的Mirage软件进行容积重建。四维超声心动图使用新型旋转扫描头原型在心动周期中采集超声图像,用于重建作为时间函数的容积变形。使用门控SPECT和四维超声心动图测量舒张末期容积、收缩末期容积和射血分数。
获得的结果显示两种方法测量的容积之间具有良好的相关性,但门控SPECT测量的容积略有高估。为了校正高估,对门控SPECT测量容积时滤波和像素大小参数的影响进行了量化。
校正后的门控SPECT与四维超声心动图之间的一致性证实了比较的相关性。本研究是在进行临床试验以全面比较两种方法获得的左心室容积之前的初步步骤。